1-4 Points to be careful about in handling.
WHEN TOUCHING LOAD AFTER TURNING OFF HIGH VOLTAGE
1.
Make the setting of an output voltage to zero (0). Turn off the HV ON/OFF switch.
2.
Check and confirm that the voltage is zero at an output voltmeter of this unit.
3.
Turn off the POWER ON/OFF switch.
4.
Earthling an output for more than 10 seconds, check and confirm that the voltage is
zero with another high voltage meter. It is especially dangerous that the load is
capacitive or a long cable is used.
5.
Make it a rule to touch load with right hand.
How to GROUND
・
For the safe operation, be sure to ground the ground terminal of power supply at
one point.
If the Load frame is not grounded
If the Load frame is grounded
・Make sure to ground properly as per the above sketch Insufficient ground can cause electrical shock
or serious damage to the unit.
・In case output short circuit or arcing is expected, please make the ground wire shorter and thicker.
1 Introduction
output cable
ground terminal
ground
load frame
load
Wires should be :
・
more than 5SQ
・
less than about 2m
・
Not bared wire
Ground at one point
rack
–
mount case
output cable
ground terminal
load frame
load
Wires should be :
・
more than 5SQ
・
less than about 2m
・
Not bared wire
Ground at one point
rack
–
mount case