20
AmpTip
TM
clamp-on DCA
RANGE
Accuracy
1) 2) 3)
60.00A
2.0% + 5d
1) Induced error from adjacent current-carrying conductor: <0.01A/A
2) Specified with DC-Zero mode applied to offset the non-zero residual
readings, if any
3) Add 10d to the specified accuracy @ < 4A
AmpTip
TM
clamp-on DC+ACA
RANGE
Accuracy
1) 2) 3)
DC, 50Hz ~ 60Hz
60.00A
2.0% + 7d
1) Induced error from adjacent current-carrying conductor: <0.01A/A
2) Specified with DC-Zero mode applied to offset the non-zero residual
readings, if any
3) Add 10d to the specified accuracy @ < 4A
Regular Clamp-on ACA
RANGE
Accuracy
1) 2) 3)
50Hz ~ 100Hz
60.00A
4) 5)
, 600.0A
1.8% + 5d
100Hz ~ 400Hz
60.00A
4) 5)
, 600.0A
2.0% + 5d
1) Induced error from adjacent current-carrying conductor: <0.01A/A
2) Maximum Crest Factor < 2 : 1 at full scale & < 4 : 1 at half scale
3) Add 10d to the specified accuracy @ < 9A
Regular Clamp-on DCA
RANGE
Accuracy
1) 2)
60.00A
3)
, 600.0A
2.0% + 5d
1) Induced error from adjacent current-carrying conductor: <0.01A/A
2) Specified with DC-Zero mode applied to offset the non-zero residual
readings, if any
3) Add 10d to the specified accuracy @ < 9A