3-5
Masterflex REGLO Digital Pump Drive with Advanced Connectivity
Section 2: Basic Setup & Settings
Masterflex
Miniflex Models
CAUTION
: Turn the drive off before removing or installing tubing. Fingers or loose clothing could
get caught in the drive mechanism.
• Check to ensure that the rollers are clean and free of defects.
• Tubing should be clean and routed so that bend radii are at a minimum of four (4) times the outside tube
diameter and are as short as possible.
• Use the appropriate tube diameter for the required flow rate and viscosity.
• Regularly recalibrate tubing to maintain the best accuracy of flow rates (for further information see "
• After installing new tubing the pump should be run for 20-30 minutes to properly condition the tubing
and increase dispense accuracy (for further information see "
to condition new tubing
" below).
1. Turn the pump head handle counterclockwise to raise the upper section of the pump head.
2. Place the tube into the opening, ensuring that the tube is:
a. Completely inside the pump head.
b. Seated over the roller-head.
c. Sitting in the V-shaped grooves located on either side of the pump head.
3. Turn the pump head handle clockwise to close the pump head.
STEP 1
STEP 2
STEP 3