5. Measuring Operations

5.1 AC Voltage and DC Voltage Measurement

1.To avoid electricial shock or meter damage, any 
    voltage in excess of 600V DC or r.m.s AC are not 
    allowed to measure.
2.To avoid electricial shock or meter damage, any 
    voltage in excess of 600V DC or r.m.s. AC 
    between meter and earth aren't allowed to apply.
3.Don't measure AC/DC voltages if a motor on the 
   circuit is being switched ON or OFF. Large voltage 
   surges may occur that can damage meter.

Warning

The voltage ranges of the meter are:
DC Voltage: 400.0mV, 4.000V, 40.00V, 400.0V, 600V;
AC Voltage: 4.000V, 40.00V, 400.0V, 400.0V.600V

The method of measuring voltage is:

1.Turn rotary switch to       

position.

Hz% 

2.Press        key and select AC or DC measuring Apply 
   the two ends of test leads to measure the voltage 
   value of circuit under test. (Paralleled with circuit 
   under test).
3.Read measured voltage value with LCD . When 
   measuring DC voltage ,the display shall display the 
   voltage polarity connected with red test leads at the 
   same time.

Notice

:

At DC 400mV and AC 4V ranges and under non-input, 
the meter shall have several readings because of 
outside interference, but its normal usage and 
measuring accuracy will not be affected.

5.2 Resistance Measurement

In order to avoid the meter or measured equipment
from being damaged,cut off all power supply of 
measured circuits and discharge all high voltage
capacitors before measuring capacitance.

Warning

The

 

resistance

 

ranges

 

of

 

the

 

meter

 

are:

400.0

Ω

,4.000k

Ω

,40.00

 

k

Ω

,400.0

 

k

Ω

,4.000M

Ω

,40.00

 

M

Ω

.

The method of measuring resistance is:

1.Turn rotary switch to        position.
2.Press        key and select 

Ω 

measuring range.

3.Apply the two ends of test leads to measure the 
   resistance value of circuit under test.
4.Read measured resistance value with LCD.

Notice:

1.Theno-line measured resistance value shall be 
    different from rated resistance value, because other 
    elements on circuit are connected with measured 
    resistance, which is equivalent to parallel connection 
   of two or multi resistances.
2.When measuring low resistance, for the measurement 
   accuracy, please firstly make two test leads short 
   circuit , here read out the displayed resistance value, 
   minus this displayed value from the correct counting of 
   measured resistance.
3.When measuring at high resistance range, the reading 
   can be confirmed after several seconds.
4.Under open circuit , if the meter displays “OL”, it will 
   show that the measured value exceeds range.

Summary of Contents for MS8332C

Page 1: ...MS8332C Digital Multimeter User Manual CAT III 600 V...

Page 2: ...5 4 2 Electrical Technical Index 5 5 Measuring Operations 9 5 2 Resistance Measurement 10 5 1 AC Voltage and DC Voltage Measurement 9 5 3 Diode Measurement 11 5 4 Continuity Measurement 11 5 5 Capacit...

Page 3: ...The special attention should be paid when using the meter because the improper usage may cause electric shock and damage the meter The safety measures in common safety regulations and operating instr...

Page 4: ...o hold the data Press this key 2 second to open the backlight 3 3 Symbols 3 Product Overview 3 1 Product External View 1 Non contact AC Voltage Indicator Light 2 LCD Display 3 HOLD backlight Key 4 Fun...

Page 5: ...32 104 relative humidity 80 RH Storage temperature and humidity 10 50 14 122 relative humidity 70 RH Altitude 2000m Power Supply AAA 2 1 5V Battery Safety Class EN61010 1 CAT III 600V Dimension LxWxH...

Page 6: ...y 4000uA 40mA 400mA 10uA 1uA 100uA 400uA 0 1uA Max Current 400mA Overload Protection mA measuring range FF 400mA H 600V 10KA fuse Frequency Range 40 400Hz Response Average value r m s of sine wave 2 0...

Page 7: ...f outside interference but its normal usage and measuring accuracy will not be affected 5 2 Resistance Measurement In order to avoid the meter or measured equipment from being damaged cut off all powe...

Page 8: ...ing The method of continuity measurement is 1 Turn rotary switch to position 2 Press key and select measuring range 3 Separately connect two ends of test leads to the two ends of measured part and cir...

Page 9: ...suring resistance is 1 Cut off the power of measured circuit Discharge all high voltage capacitance on measured circuit 2 Turn rotary switch to mA or uA position When the measured current is less than...

Page 10: ...H 600V 10KA Don t use the instrument until the rear cover is placed back and the screws are tightened Clean the housing of instrument only with a wet rag dripped with little detergent but never chemic...

Reviews: