4.2.6 DC Current

Resolution

Accuracy

200mA

0.1mA

10mA

10A

Measuring range

±(0.8% of r3 digits)

±(1.2% of r3 digits)

If measured resistance 
is less than 50

Ω

buzzer will sound

4.2.5 Continuity 

Function

Description

Description

Open circuit 
voltage: 2.9V

Overload protection: 250V DC or AC (RMS)

Forward DC current: 1mA
Reverse DC voltage: 2.9V
Display shows forward 
voltage drop

4.2.4 Diode Test

Function

Description

Resolution

0.001V

Diode Test

Overload protection: 250V DC or AC (RMS)

12

13

4.2.7 AC Current

Resolution

Accuracy

200mA

0.1mA

10mA

10A

Measuring range

±(1

.

0% of r3 digits)

±(1.5% of r3 digits)

Overload protection: 
mA jack: FF400mA H 600V fuse
10A jack: FF10A H 600V fuse
Max input current: mA jack: 200mA DC 
10A jack: 10A  DC 
When measuring current exceeding 2A, do not measure 
for longer than 2 minutes continuously.Wait 10 minutes 
to continue measurement.

4.2.8

 

Battery Test

Resolution

Accuracy

12V

0.01V

0.01V

9V

Position

±(0.8% of r7 digits)

±(0.8% of r7digits)

1.5V

0.001V

±(3.0% of r5digits)

Overload protection: F400mA H 600V fuse

4.2.9

 

NCV Test

Overload protection: 
mA jack: FF400mA H 600V fuse
10A jack: FF10A H 600V fuse
Max input current: mA jack: 200mA AC rms
10A jack: 10A  AC rms
When measuring current exceeding 2A, do not measure 
for longer than 2 minutes continuously.Wait 10 minutes 
to continue measurement.

• Hold down the NCV button and move the NCV sensor 
  at the top of the unit next to the object being tested.
• When voltage is detected (>110V AC rms) the buzzer 
  will sound and the NCV indicator will flash.

Summary of Contents for MS8238B

Page 1: ...gital Multimeter CAT III 600V MS8238B DIGITAL MULTIMETER HOLD NCV 200 600 200m 200m 200 2k 20k 200k 2M 2 20 200 600 200m 9V 1 5V 12V A V V A 600V CAT III mAV COM MAX 30 sec every 15 min MAX FUSED 10A...

Page 2: ...mbols on the meter are to advise of potential dangerous situations Caution is required when measuring close to the mete s safety limits Never exceed the protection limit values indicated in the specif...

Page 3: ...ower and check fused before connect the leads when measuring current Turn circuit power on after making connection Never use the meter unless the back cover is in place and fastened securely When the...

Page 4: ...ment display 4 Polarity indicator 3 4 1 2 8 1 2 7 3 4 5 6 6 Input jack all functions except current greater than 200mA 7 Non contact voltage NCV button 8 Non contact voltage NCV indicator MS8238B DIGI...

Page 5: ...current range Move the rotary switch to a higher range Set the rotary switch to the appropriate AC or DC Warning Do not measure voltages higher than 600V DC or AC rms to perevent damage to the meter o...

Page 6: ...red test lead to the input jack and the black lead to the COM jack Connect the leads to the circuit under test If the measured resistance is less than 50 the buzzer will sound Warning To prevent inju...

Page 7: ...LCD display Overload Indication Display shows OL Low Battery Indication 4 2 Technical Specifications Accuracy of reading digits at 18 C 28 C with a relative humidity of 80 guaranteed for a period of...

Page 8: ...k FF10A H 600V fuse Max input current mA jack 200mA DC 10A jack 10A DC When measuring current exceeding 2A do not measure for longer than 2 minutes continuously Wait 10 minutes to continue measurement...

Page 9: ...ection could be impaired by socket design insulation etc When measuring voltage through input jacks NCV indicator may light up due to induced voltage External environmental interference could cause fa...

Page 10: ...R 00 05 1049...

Reviews: