background image

37633/8722A-0115 

 

 

 

 
 
 
MightySat

 

Fingertip Pulse

 

Oximeter 

 

  Manufacturer: 

 

 

Masimo Corporation 

40 Parker 

Irvine, CA 92618 

USA 

Tel.: 949-297-7000 

Fax.: 949-297-7001 

www.masimo.com 

 

 

MightySat Fingertip Pulse Oximeter 

Operator's Manual 

For Sale in the USA- Not Intended for Medical Use 

 

Notice: Purchase or possession of this device does not carry any express or implied license to use with 

replacement parts which would, alone or in combination with this device, fall within the scope of one of the 

relating patents. 

 

About this Manual 

Do not operate the MightySat Fingertip Pulse Oximeter without completely reading and understanding the 

instructions. 
Always use the MightySat precisely in accordance with the directions in this manual, including site 

selection and sensor placement. Failure to follow all of the directions in this manual could lead to 

inaccurate measurements.   
Read and follow any warnings, cautions, and notes presented throughout this manual. The following are 

explanations of warnings, cautions, and notes. 
A warning is given when actions may result in a serious outcome to the user (for example, injury, serious 

adverse effect, or death). 

Warning:

 This is an example of a warning statement. 

A caution is given when any special care is to be exercised by the user to avoid injury to the user, damage to 

this instrument, or damage to other property. 

Caution:

 This is an example of a caution statement. 

A note is given when additional general information is applicable. 

Note:

 This is an example of a note. 

 

Product Description and Intended Use 

Product Description 

The MightySat Fingertip Pulse Oximeter is a noninvasive device that measures and displays arterial oxygen 

saturation (SpO2), Pulse Rate (PR), Perfusion Index (PI), and Pleth Variability Index (PVI®). The MightySat 

is available for people who weigh more than 30 kg (66 lbs). 
The MightySat Fingertip Pulse Oximeter is available in the following models: 

Model 

Product   

Features 

9700 

MightySat 

Measures and displays arterial oxygen saturation (SpO2), Pulse Rate (PR), and 

Perfusion Index (PI). 

9800 

MightySat, Bluetooth 

Measures and displays arterial oxygen saturation (SpO2), Pulse Rate (PR), and 

Perfusion Index (PI). 
Bluetooth LE radio for transfer of parameter data to a compatible smart device. 

9900 

MightySat, Bluetooth and PVI 

Measures and displays arterial oxygen saturation (SpO2), Pulse Rate (PR), 

Perfusion Index (PI), and Pleth Variability Index (PVI).   
Bluetooth radio for transfer of parameter data to a smart device. 

 

Intended Use 

The MightySat Fingertip Pulse Oximeter by Masimo is for use by sports and aviation users who are 

interested in knowing their blood oxygenation level (SpO2), pulse rate (PR) and perfusion index (PI). The 

MightySat is not intended for medical use. The MightySat can be used in a wide range of settings, including 

extreme sports, and for passengers and pilots. 

 

Safety Information, Warnings, and Cautions 

Safety Warnings and Cautions 

 

Warning:

 Do not use MightySat if it appears or is suspected to be damaged. 

 

Warning:

 Do not repair, open, or modify MightySat. Injury or equipment damage could occur. 

 

Warning:

 Do not use MightySat during magnetic resonance imaging (MRI) or in an MRI 

environment. 

 

Warning:

 Do not expose MightySat components to excessive moisture such as direct exposure to 

rain. Excessive moisture can cause the components to perform inaccurately or fail. 

 

Warning:

 Do not place MightySat or accessories in any position that might cause it to fall on the 

user. 

 

Warning:

 Electric Shock Hazard: Do not use MightySat during defibrillation. 

 

Warning:

 Do not to use device during electrosurgery. 

 

Warning:

 Do not use MightySat in the presence of flammable anesthetics or other flammable 

substance in combination with air, oxygen-enriched environments, or nitrous oxide. 

 

Warning:

 Do not use tape to secure MightySat to the site; this can restrict blood flow and cause 

inaccurate readings. Use of tape can cause skin damage, and/or pressure necrosis or damage the 

sensor. 

 

Warning:

 Keep the oximeter away from young children. Small items such as the battery door and 

battery may become choking hazards. 

 

Warning: 

When MightySat is applied for extended periods of time, the sensor site must be checked 

frequently to ensure adequate circulation, skin integrity, and correct optical alignment. 

 

Warning: 

If MightySat is applied too tightly or becomes tight due to edema it will cause inaccurate 

readings and can cause pressure necrosis. 

 

Warning:

 Do not use the lanyard with MightySat during activities where a risk of strangulation may 

exist (i.e. sleeping). 

 

Caution:

 Do not attempt to reprocess, recondition or recycle MightySat as these processes may 

damage the electrical components, potentially leading to user harm. 

 

Caution: 

Before changing batteries, make sure the device is off and is not applied to a finger. 

 

Caution:

 MightySat contains a magnet which may interfere with other devices, including 

pacemakers. 

 

 

Note:

 The maximum skin surface temperature is measured to be less than 41°C (106°F) in a 

minimum 35°C (95°F) environment. This was verified by measuring the skin interface temperature 

with MightySat operating under reasonable worst-case conditions.   

 

Performance Warnings and Cautions 

 

Warning

: MightySat is not intended for medical use. 

 

Warning

: MightySat is not an apnea monitor and should not be used for arrhythmia analysis. 

 

Warning:

 Do not self-diagnose or self-medicate on the basis of the measurements. Always consult 

your doctor. MightySat is not intended for medical use. 

 

Warning:

 MightySat does not include visual alerts or audible alarm features to support continuous 

monitoring. It is intended for spot-check use only. 

 

Warning: 

SpO2 is empirically calibrated in healthy adult volunteers with normal levels of 

carboxyhemoglobin (COHb) and methemoglobin (MetHb). 

 

Warning: 

Inaccurate SpO2 readings may be caused by: 

 

Improper MightySat placement or alignment 

 

Elevated levels of COHb and MetHb: High levels of COHb or MetHb may occur with a 

seemingly normal SpO2. When elevated levels of COHb or MetHb are suspected, laboratory 

analysis (CO-Oximetry) of a blood sample should be performed. 

 

Intravascular dyes such as indocyanine green or methylene blue 

 

Externally applied coloring and texture such as nail polish, acrylic nails, glitter, etc. 

 

Users attached to a high pressure cuff 

 

Avoid placing the MightySat sensor on any extremity with an arterial catheter or blood 

pressure cuff. 

 

Elevated levels of bilirubin 

 

Severe anemia 

 

Venous congestion   

 

Venous pulsation 

 

Low arterial perfusion 

 

Motion artifact 

 

Caution:

 To minimize radio interference, other electrical equipment that emits radio frequency 

transmissions should not be in close proximity to MightySat. 

 

Caution:

 High ambient light sources, fluorescent lights, infrared heating lamps and direct sunlight 

can interfere with the performance of MightySat. 

 

Caution:

 To prevent interference from ambient light, ensure that MightySat is properly applied. 

Failure to take this precaution in high ambient light conditions may result in inaccurate or no 

measurements. 

 

Caution:

 If no measurements are displayed, find a better perfused monitoring site. 

 

Caution: 

When using

 

MightySat Model 9800 or Model 9900 with a smart device, keep both devices 

within the recommended range of each other (see 

Specifications

 for details); moving outside of this 

range may cause missing, lost, and/or inaccurate data to display on the smart device. 

 

Note:

The MightySat display may be difficult to view when exposed to direct sunlight or bright lights. 

 

Note:

 A functional tester cannot be used to assess accuracy of MightySat. 

 

Note:

 The MightySat display will shut off automatically if there are no readings. 

 

Cleaning and Service Warnings and Cautions 

 

Warning:

 Changes or modifications not expressly approved by Masimo shall void the warranty for 

this equipment and could void the user’s authority to operate the equipment. 

 

Warning:

 Use only AAA alkaline batteries. 

 

Warning:

 Batteries may leak or explode if used or disposed of improperly. Remove batteries if the 

device will be stored for more than 30 days.   

 

Warning: 

Do not mix fully charged and partially charged batteries at the same time. These actions 

may cause the batteries to leak. 

 

Caution:

 A user may only perform maintenance procedures specifically described in the manual; 

otherwise, return MightySat for servicing. 

 

Caution:

 Do not use petroleum-based or acetone solutions, or other harsh solvents, to clean 

MightySat. These substances affect the device’s materials and instrument failure can result. 

 

Caution:

 Do not submerge MightySat in any cleaning solution or attempt to sterilize by autoclave, 

irradiation, steam, gas, ethylene oxide or any other method. This will seriously damage the device. 

 

Caution:

 Do not use undiluted bleach (5% - 5.25% sodium hypochlorite) or any cleaning solution 

other than those recommended here because permanent damage to MightySat may occur. 

 

Caution: 

Never immerse MightySat in water or any other liquid solution.   

 

Compliance Warnings and Cautions 

 

Caution:

 Disposal of Product: Comply with local laws in the disposal of the instrument and/or its 

accessories, including batteries. 

 

Note:

 When using MightySat with a device with wireless features, consideration should be taken to 

local government frequency allocations and technical parameters to minimize the possibility of 

interference to/from other wireless devices.   

 

Note:

 In accordance with international telecommunication requirements, the frequency band of 2.4 

GHz is only for indoor usage to reduce potential for harmful interference to co-channel mobile 

satellite systems. 

 

Note: 

This device complies with Part 15 of the FCC Rules. Operation is subject to the following two 

conditions: (1) This device may not cause harmful interference, and (2) this device must accept any 

interference received, including interference that may cause undesired operation. 

 

Note:

 This equipment has been tested and found to comply with the limits for a Class B digital 

device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable 

protection against harmful interference in a residential installation. This equipment generates, uses, 

and can radiate radio frequency energy and, if not installed and used in accordance with the 

instructions, may cause harmful interference to radio communications. However, there is no 

guarantee that interference will not occur in a particular installation. If this equipment does cause 

harmful interference to radio or television reception, which can be determined by turning the 

equipment off and on, the user is encouraged to try to correct the interference by one or more of the 

following measures:   

 

Reorient or relocate the receiving antenna. 

 

Increase the separation between the equipment and receiver. 

 

Connect the equipment into an outlet on a circuit different from that to which the receiver is 

connected. 

 

Consult the dealer or an experienced radio/TV technician for help. 

 

Note

: This equipment has been tested and found to comply with the Class B limits for medical 

devices according to the IEC 60601-1-2:2007. These limits are designed to provide reasonable 

protection against harmful interference in all establishments, including domestic establishments. 

 

The Masimo Difference 

From the leader in hospital pulse oximetry comes a complete pulse oximeter that is compact and portable 

to help you measure blood oxygen saturation through motion and low perfusion. The MightySat Fingertip 

Pulse Oximeter allows you to noninvasively track oxygen saturation (SpO2), pulse rate (PR), perfusion index 

(PI), and Pleth Variability Index (PVI). The MightySat features Masimo® Signal Extraction Technology® 

(SET®) technology, proven to provide accurate pulse oximetry measurements during motion and low 

perfusion. 
For more than 20 years, Masimo has been focused on a singular mission–to take noninvasive patient 

monitoring to new sites and applications. Visit www.Masimo.com for more information about Masimo 

products. 

 

About the MightySat Technology 

Pulse oximetry is governed by the following principles: 
Oxyhemoglobin (oxygenated blood) and deoxyhemoglobin (non-oxygenated blood) differ in their 

absorption of red and infrared light (spectrophotometry). 

Display Parameter 

Description 

Functional oxygen saturation (SpO2) 

Functional oxygen saturation (SpO2) is the amount of oxyhemoglobin expressed as a 

percentage of the hemoglobin that is available to transport oxygen.   

Pulse Rate (PR) 

Pulse rate (PR), measured in beats per minute (BPM), is based on the optical detection of 

peripheral flow pulse. 

Perfusion Index (PI) 

The Perfusion Index (PI) is the ratio of the pulsatile blood flow to the non-pulsatile or static 

blood in peripheral tissue. PI thus represents a noninvasive measure of peripheral perfusion 

that can be continuously and noninvasively obtained from a pulse oximeter. 

Pleth Variability Index (PVI ) (Model 9900 

only) 

The Pleth Variability Index (PVI) is a measure of the dynamic changes in the Perfusion Index 

(PI) that occur during the respiratory cycle. The calculation is accomplished by measuring 

changes in PI over a time interval where one or more complete respiratory cycles have occurred. 

PVI is displayed as a percentage (0-100%). 
The utility of PVI is unknown at this time and requires further clinical studies. Technical factors 

that may affect PVI include malposition and user motion. 

Plethysmograph Waveform (Pleth) 

The amount of arterial blood in tissue changes with your pulse (photoplethysmography). 

Therefore, the amount of light absorbed by the varying quantities of arterial blood changes as 

well. Clinicians have attempted to utilize this waveform and its variation for assigning signal 

integrity, physiological and artifactual changes such as perfusion changes, dysrhythmia, motion 

artifact, and electrical interference. For this reason, accurate and reliable presentation of the 

plethysmograph waveform is of importance. 

Signal I.Q. Indicators 
(SIQ) 

The Signal I.Q.® (SIQ) provides an indicator of the assessment of the confidence in the 

displayed SpO2 value. The height of the vertical line of the SpO2 SIQ provides an assessment of 

the confidence in the measurement displayed. A high vertical line indicates higher confidence 

in the measurement. A small vertical line indicates lower confidence in the displayed 

measurement.   

 

 