Ref. No: m94B/OM/101 
Issue No: 00

 

 

Operator’s manual 

Page 10 of 16 

 

 

HT16u unit equipped with 2 push button key labeled as Menu key and Up key. 
 

UP Key 

 

Menu Key 
 

(1)

MIn

 Mode: 

It displays minimum value of measured 

Temperature and Humidity 

from last records.

 

(2)

MAX

 Mode: 

It displays maximum value of measured 

Temperature and Humidity 

from last records.

 

(3)

TSPL

 (Temperature Set point Low) Mode: 

TSPL

 stands for 

Temperature Set Point Low. It is user input menu, user can simply 

enter 2 Digit Temperature Set Point Low value followed by 1 decimal point value with 
the help of menu key & up key. If Actual Temperature goes low below TSPL set value 
then it will send Temperature low alarm indication on LCD as small ‘L’. Temperature 
Low Alarm indication appear as symbol ‘

L

’ ahead of temperature reading during run-

mode. 

(4)

TSPh

 (Temperature Set Point High) Mode: 

TSPh

 stands for 

Temperature Set Point High. It is user input menu, user can simply 

enter 2 Digit Temperature Set Point High value followed by 1 decimal point value with 
the help of menu key & up key. If Actual Temperature goes high above TSPH set value 
then it will send Temperature High alarm indication on LCD as small ‘u’. Temperature 
High Alarm indication appear as symbol ‘

v

’ ahead of temperature reading during run-

mode. 

(5)

hSPL

 (Humidity Set Point Low) Mode: 

hSPL

 stands for 

Humidity Set Point Low. It is user input menu, user can simply enter 2 Digit 

Humidity Set Point Low value followed by 1 decimal point value with the help of menu key & 
up  key.  If  Actual  Humidity  goes  low  below  HSPL  set  value  then  it  will  send  Humidity  low 
alarm  indication  on  LCD as  small  ‘L’.  Humidity  Low  Alarm  indication  appear  as  symbol ‘

L

’ 

ahead of humidity reading during run-mode.

 

(6)

hSPh

 (Humidity Set Point High) Mode:

 

hSPh 

stands for 

Humidity Set Point High. It is user input menu, user can simply enter 2 Digit 

Humidity Set Point High value followed by 1 decimal point value with the help of menu key 
& up key. If Actual Humidity goes high above HSPH set value then it will send Humidity high 
alarm indication on LCD as small ‘u’. Humidity High Alarm indication appear as symbol ‘

v

’ 

ahead of humidity reading during run-mode. 

(7)

dd.mm.yyyy

 (Date) Mode: 

This  mode  shows  the  current  date,  month  &  year  of  data-logger.  User  can  set  date, 
month & year via SMART-HT configuration software.   

Summary of Contents for HT16U

Page 1: ...User Manual HT16U HUMIDITY TEMPERATURE SMARTLOGGER DOC m94B OM 101Issue No 00 Copyright 2018 Masibus Automation and Instrumentation Pvt Ltd Customer Support support masibus com Toll Free Number India...

Page 2: ...6 HT16u Configuration 7 6 1Configuration via front key Panel 7 6 2 Main menu Flowchart 8 6 3 Main menu description 9 6 4 Quick Log Active Menu Flowchart 12 6 5 LCD Messages 13 1 Logging START Indicati...

Page 3: ...mes extremely low power operates on two Lithium Coin cell is housed in an elegant compact wall mount enclosure and has two front panel keys for operation The Humidity Temperature Logger has a micro US...

Page 4: ...ax Min Alarm status Keys Two for operation Data Logging Logging Start Stop Manual or pre programmed with pause facility Memory 64MB Record Fields Date Time Temperature Humidity Alarm Status Log Interv...

Page 5: ...Ref No m94B OM 101 Issue No 00 Operator s manual Page 5 of 16 5 HT16u MOUNTING...

Page 6: ...Ref No m94B OM 101 Issue No 00 Operator s manual Page 6 of 16...

Page 7: ...e Mode 3 TSPL Temperature Set Point Low Mode 4 TSPh Temperature Set Point High Mode 5 hSPL Humidity Set Point Low Mode 6 hSPh Humidity Set Point High Mode 7 dd mm yyyy Date Mode 8 hh mm ss Time Mode 9...

Page 8: ...Ref No m94B OM 101 Issue No 00 Operator s manual Page 8 of 16 6 2 Main menu Flowchart...

Page 9: ...n the device unit will display Temperature and Humidity measured by Sensor Device also shows battery level alarm indication Log START STOP Status When device stay inactive or user don t press any key...

Page 10: ...tual Temperature goes high above TSPH set value then it will send Temperature High alarm indication on LCD as small u Temperature High Alarm indication appear as symbol v ahead of temperature reading...

Page 11: ...isplay status change to fdsp off and display update at interval of LTMR value record interval 11 mode Sensor Type Mode This mode is for sensor selection Sensor type of data logger can be change with t...

Page 12: ...e mark for important event indication Mark event would be written in logger every time mark added User can enter in this menu quickly with following key combination As shown in figure user can enter i...

Page 13: ...ted When data logging start by key panel or SMART HT configuration software unit will punch event log as START in data report 2 Logging PAUSE Indication Example below represent Logging PAUSE indicatio...

Page 14: ...state When data logging stop by key panel or SMART HT configuration software unit will punch event log as STOP in data report 4 USB Connection Indication Example below represents USB communication ind...

Page 15: ...ct such as a flat tip screw driver or tweezers to lift the battery from the compartment 4 Insert the new battery with the positive polarity facing up at an angle under the metal latch Push the end in...

Page 16: ...al Page 16 of 16 CONTACTDETAILOFSERVICEDEPARTMENT Masibus Automation Instrumentation Pvt Ltd B 30 GIDC Electronics Estate Sector 25 Gandhinagar 382044 Gujarat India 91 79 23287275 79 91 79 23287281 82...

Reviews: