The problem can be bad connection / power of incorrect
rating.
First check, power on terminal of the instrument; if it is not
present, then the fault is in power chord.
One must take care while dealing with Power
wirings because it may create electrical shock.
2.
UNSTABLE / Vague READING
Check for loose connections.
The reason can be reverse input connections
3.
OUTPUT
NOT
MATCHING
WITH
THE
EXPECTED VALUE
It is a normal tendency to doubt the instrument performance,
when the Output is not matching the expected value. Kindly
make sure that the output is really incorrect with respect to
input signal, before attempting any re-calibration.
Account for measuring instrument’s inaccuracies, lead errors
and calibration errors. Care must be taken when measuring
Output signal.
Use calibrating instrument of accuracy better than 0.1% for
purpose of calibration. If the signal is still found to be out of
tolerance, calibration should be required.