3. Putting Into Operation
3.1. Opening the Instrument
WARNING!
Before opening the instrument disconnect the power cord from the power supply.
To open the instrument unscrew the 8 screws from the back and lift the cover.
3.2. Connecting a Weighing Platform
●
Insert the weighing platform cable into the instrument through the cable gland
●
Connect the weighing platform cable to J3 terminal according to the
table below.
Pin
1
2
3
4
5
6
Assignment
SIG+
SIG-
EXC+
EXC-
REF+
REF-
Note:
For 4 wire loadcells J1 and J2 pads should be solder bridged or Ref+ and Ref- shoud be connected to EX+
and EX- respectively.