MAS 640 Instruction Manual Rev. 0A
8. Saving Changes
In the main menu setup, press the <PIECES> key repeatedly until 'SauE' is
displayed.
< PIECES>
Accept the savings by pressing the <FUNCTION> key.
<FUNCTION>
9. Leaving Setup mode
After completion of all the adjustments select the 'SAuE' function from the main menu and accept it by pressing
the <FUNCTION> key.
●
Make sure the calibration jumper is in the position J10.
●
Power cycle the instrument.
Warning:
New settings will not be stored unless saved in non volatile memory
10. Error Messages Table
Error:
Description:
Solution:
-E06-
The readings are not stable.
Retry an operation when the readings
become stable.
-E07-
The initial load exceeds 10% of the
maximum range.
Remove any load from the platform before
turning on the instrument.
-E08-
Counting pieces resolution error occurred. For the counting mode, mass of the single
piece should represent at least 1/4 of the
instrument minimum displayed division. Use
heavier item for counting.
-E02-
Load exceeds the maximum measuring
range of the instrument.
Remove heavy load from the platform.
-E01-
Incorrect tare value occured.
Taring is only allowed for indication higher
than zero.
Use the<Tare> key to clear the previous tare
while indication is below or equal to zero.
-19-