background image

2

2. PRODUCT SPECIFIC SAFETY INFORMATION

Measurement Category IV (CAT IV)

 is applicable to test and 

measuring equipment connected at the source of the building’s low-
voltage MAINS installation.

The specifi ed measurement category means the phase sequence 
indicator will be safe to the user if inadvertently connected to a voltage 
up to 600V AC/DC to earth within a CAT IV environment. It 

does 

not 

mean it can be used to test for a voltage beyond its maximum 

specifi ed limits.

 
2.1 Precautions

This product has been designed with your safety in mind, but please 
pay attention to the following warnings and cautions before use.
 

Warning

The phase sequence indicator must only be used by a skilled and 
competent person who is familiar with the relevant regulations, 
the safety risks involved and the consequent normal safe working 
practices.

Before use check the unit for cracks or any other damage. The cables 
have black outer and contrasting inner insulation, to allow damage to 
the cables to be easily identifi ed. If there is any doubt the unit should 

not be used

.

 
Always verify the unit is functioning correctly on a known correctly 
sequenced 3-phase voltage source before and after use. 

Do not use

 

the unit if the unit does not give the correct indications.

Always keep your fi ngers behind the test probe fi nger guards. Never 
touch the exposed metal probe tips.

PC15250 instructions rev3.indd   4

09/07/2013   08:25:32

3

The socket tester is not intended to determine if circuits being 
tested are dead. The correct testing methods should be used to do 
this. 

If the LED indicators do not illuminate, this does not necessarily 
mean all three phases under test are dead. Two phases could be 
open circuit, but one phase could still be live.

Caution

  

Avoid severe mechanical shock or vibration and extreme 
temperature.

PC15250 instructions rev3.indd   5

09/07/2013   08:25:32

4

OPERATION
3.1

 

Description of LED Indicators

 
                             

L1

L2

L3

THREE LIT: CORRECT    

ONE LIT: THAT PHASE FAULTY

PHASE

PHASE SEQUENCE

L1

L2

L3

L1

L2

L3

 

Both phase sequence LED’s are illuminated if a phase is missing.

3.2 Use of Test Probe Shrouds

The shrouds around the L1 and L2 test probes are normally sprung 
forwards to IP 2X rating and are retractable.

For optimum safety, the shrouds should be allowed to spring forward 
freely whenever the probe tips are removed from a location under test.

If desired, and before the unit is connected to any source of voltage, the 
shrouds can be locked back by pushing and twisting 90°.

The probe tips will be exposed by 3 ± 0.5mm. In this position they are 
GS38 compliant.

3.3 Proving Check

Before and after use, verify the unit is functioning correctly on a known 
correctly sequenced 3-phase voltage source of 50V to 600V phase to 
phase, and frequency of 40Hz to 60Hz.

All phases present when all
three LED’s illuminated

One LED illuminated 
indicates the respective 
phase is missing

Incorrect phase 
sequence of 
L1, L3, L2 when 
illuminated*

Correct phase 
sequence of 
L1, L2, L3 when 
illuminated*

PC15250 instructions rev3.indd   8

09/07/2013   08:25:33

5

Connect the L1, L2 and L3 test probes respectively to the L1, L2 and 
L3 phases of the voltage source.

The PC15250 must indicate as shown below.

 

 

 

L1, L2 & L3 LED’s must illuminate red

 

 

  

 Phase Sequence L1 L2 L3 LED must 
illuminate red

 

 

  

 Phase Sequence LED L1 L3 L2 LED 
must be extinguished

L1

L2

L3

THREE LIT: CORRECT    

ONE LIT: THAT PHASE FAULTY

PHASE

PHASE SEQUENCE

L1

L2

L3

L1

L2

L3

Do not use 

the unit if the unit does not give the correct indications 

above.

During this verifi cation emphasis should also be placed upon the 
fl exing of the individual cables along their length, and particularly at 
the entry points to the hand held elements, to confi rm that the cable/s 
have not fractured.

3.4 Testing a 3 Phase Distribution System

Take all required safety precautions when connecting the test probes 
to the 3-phase electrical distribution system to be tested. 

1.   The three test probes are marked L1, L2, L3. These relate to the 

L1, L2, L3 LED’s on the instrument facia.

2.  To determine phase presence and phase sequence, connect 

the test probes to the phase cables of the 3-phase electrical 
distribution system to be tested. Refer to table 2 of section 3.5 to 
determine to which cables to connect L1, L2 and L3 (see Note 2).

PC15250 instructions rev3.indd   9

09/07/2013   08:25:33

Reviews: