RTD_TYPE?
IEEE-488
RS-232
Sequential
Overlapped
This command returns the Resistance Temperature Detector (RTD) sensor type being
used for RTD temperature source and measurement.
Parameter:
<None>
Response:
<value>
where <value> is one of the following:
PT385_100
100-ohm RTD, curve a=0.00385 ohms/ohm/°C
PT385_200
200-ohm RTD, curve a=0.00385 ohms/ohm/°C
PT385_500
500-ohm RTD, curve a=0.00385 ohms/ohm/°C
PT385_1000
1000-ohm RTD, curve a=0.00385 ohms/ohm/°C
PT392_100
100-ohm RTD, curve a=0.003926 ohms/ohm/°C
PTJIS_100
100-ohm RTD, curve a=0.003916 ohms/ohm/°C
CU10
10-ohm RTD, empirical curve
NI120
120-ohm RTD, empirical curve
YSI_400
YSI thermistor curve
OHMS_HIGH
4000 ohms range
OHMS_LOW
400 ohms range
SPRT
standard PRT with user defined error coefficients, only
available for measurement
USR_DEF<x> RTD with user defined custom coefficients, where x is
the curve number and ranges from 1 to 5, i.e.
USR_DEF2 for curve 2
Example:
RTD_TYPE?
PTJIS_100
This example indicates that the RTD sensor type is a 100-ohm RTD
with curve a=0.3916 ohm/°C.