BAR
This example indicates that the primary pressure display units are bars
RANGE?
IEEE-488
RS-232
Sequential
Overlapped
This command returns the present DC voltage or current output range.
Parameter:
<None>
Response:
<value>
where <value> is one of the following:
V_0.1V
DC volts, 100 mV range
V_1V
DC volts, 1V range
V_10V
DC volts, 10 V range
V_100V
DC volts, 100 V range
A_0.1A
DC current
NONE
neither volts nor current is selected
Example:
RANGE?
V_10V
This example indicates that the present output range is 10 V.
RANGELCK
IEEE-488
RS-232
Sequential
Overlapped
This command locks or unlocks the DC voltage output range per the present output
value.
Parameter:
<value>
where <value> is one of the following:
ON
to lock the present voltage range
OFF
to unlock the present voltage range
Response:
<None>
Example:
RANGELCK ON
If the present DC voltage output is 5 V, this example locks the range at
10 VDC.
X
X
X
X
X
X