8. Color Calibration
Color Calibration:
•
Allow both the unit to be calibrated and the Minolta® CA-310 to warm up for a minimum of 20 minutes.
•
Attach the CA-310 color probe to the update dongle.
•
With the unit turned on, insert the update dongle into the service port at the rear of the device.
•
Use the Rotary Menu navigation control and go to:
-
Color Menu
-
Color Temp
-
Cal D65/D93 to calibrate both
-
Cal D65 to calibrate only D65
-
Cal D93 to calibrate only D93
•
Press ENTER once to select and again to confirm.
•
Follow the on-screen instructions.
Notes:
1. If color probe is not attached or you make a mistake and try to calibrate an incorrect screen, an error message will be
displayed and the screen will automatically default to the previous settings.
2. If the calibration process is interrupted, the screen settings will be corrupted and the calibration process will need to be
repeated.
9. Firmware Updates
Firmware Updates:
An optional OR-SM Service Module and internet connection are required to update the firmware.
1. Download the Orchid updater software package from the Marshall website.
2. Unzip files and save. Double-click the Orchid updater program and install new firmware.
3. Turn on the Orchid unit to be upgraded and connect to computer.
4. Insert the OR-SM module in the service port.
5. Run the Orchid update program, click update.
6. The Updater will check for available software, compare, and update if applicable.
Notes:
• The update process will take approximately 8 minutes.
• If the Orchid model is not automatically detected, the appropriate model will need to be selected to update.