Radiated RF IEC
61000-4-3
3 V/m 80MHz to 2,7
GHz
3 V/m
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey
a
,
should be less than the
compliance level in each
frequency range
b
.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the MEDICAL SCALE MBF-
6000/6010 is used exceeds the applicable RF compliance level above, the MEDICAL SCALE
MBF-6000/6010 should be observed to verify normal operation. If abnormal performance is
observed, additional measures my be necessary, such as re-orienting or relocating the
MEDICAL SCALE MBF-6000/6010.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be les than 3 V/m.
Page 22