10

 

Selection 

Description 

N Units 

Output format includes the value and unit of measure. Compression values have 
positive polarity, tension values have negative polarity. 

Numeric Only 

Output format includes the value only. Polarity same as above. 

Invert Polarity 

Compression values have negative polarity, tension values have positive polarity. 
May be selected in addition to the N Units / Numeric Only selection. 

Omit Polarity 

Both directions are formatted with positive polarity. May be selected in addition to 
the N Units / Numeric Only selection. 

 
9.2 Data Output  

Communication with Series 3 gauges is achieved through the micro USB port located along the left side 
of the housing, as shown in the illustration in the 

Power

 section. Communication is possible only when 

the gauge is in the main operating screen (i.e. not in a menu or configuration area).  
 
The current reading is transmitted from the gauge when the 

DATA

 key is pressed. The gauge may also 

be controlled by an external device through the USB cha

nnel by sending ASCII command ‘?’ (no quotes). 

The command must be terminated by a CR (Carriage Return) character, 0x0D, or a CR-LF (Carriage 
Return 

 Line Feed) pair, where the Line Feed, 0x0A, is ignored.  

 

9.3 Command Responses  

In response to the readi

ng request command ‘?’ the gauge will return a string with the load data, followed 

by a space, then the load unit (if enabled under the 

USB Settings

 

 

Data format

 sub-menu). It will be 

terminated by a CR-LF pair. 
 

Example return strings:  

 

-18.78 lbF<CR><LF>     18.78 lbF of tension force 
1.724 N<CR><LF>         1.724 N of compression force    

 

The number of digits after the decimal point is dependent of the gauge’s c

apacity and resolution. 

 
By default, the minus sign (-) indicates tension, and the absence of a sign indicates compression. 
However, positive / negative polarity may be inverted or omitted under the 

USB Settings 

 Data Format

 

sub-menu.  
 
Any detected errors are reported back by means of error code *10 (illegal command). 
 
The gauge can send individual data points or continuous data to 

MESUR

®

gauge

 and 

MESUR

® 

Lite

 

software.

  

 

 

10   CALIBRATION  

 

10.1 Initial Physical Setup 

The gauge should be mounted vertically to a test stand or fixture rugged enough to withstand a load 
equal to the full capacity of the instrument. Certified deadweights or master load cells should be used, 
along with appropriate mounting brackets and fixtures. Caution should be taken while handling such 
equipment.  

 
10.2 Calibration Procedure 

 

1.  Select 

Calibration

 from the menu. The display appears as follows: 

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Summary of Contents for EK3-100

Page 1: ...Series 3 DIGITAL FORCE GAUGES User s Guide www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 2: ...y and industrial environments This User s Guide provides setup safety and operation instructions Dimensions and specifications are also provided For additional information or answers to your questions...

Page 3: ...s that should not be used with the gauge include potentially flammable substances or products items that can shatter in an unsafe manner and any other components that can present an exceedingly hazard...

Page 4: ...n icon appears in the lower left corner of the display as follows If the AC adapter is not plugged in battery power drainage is denoted in a five step process 1 When battery life is greater than 75 th...

Page 5: ...he housing An additional two holes are supplied for metric screws These holes are designed to accommodate screws in order to hold the gauge in place Mark 10 test stands include a set of thumb screws f...

Page 6: ...sing ZERO or by powering the gauge off and on 3 Primary reading The current displayed force reading See Operating Modes section for details 4 Load bar Analog indicator to help identify when an overloa...

Page 7: ...primary reading and peaks UP Navigates up through the menu and sub menus MENU Enters the main menu ESCAPE Reverts one step backwards through the menu hierarchy MODE Toggles between measurement modes D...

Page 8: ...peak compression reading observed If the actual force decreases from the peak value the peak will still be retained in the primary reading area of the display Pressing ZERO will reset the value 5 3 Pe...

Page 9: ...mpared to these limits 8 2 Configuration To configure set points select Set Points from the menu The screen appears as follows Either one two or none of the set points may be enabled To toggle between...

Page 10: ...ettings select USB Settings from the menu The screen appears as follows Communication settings are permanently set to the following Data Bits 8 Stop Bits 1 Parity None Other settings are configured as...

Page 11: ...r where the Line Feed 0x0A is ignored 9 3 Command Responses In response to the reading request command the gauge will return a string with the load data followed by a space then the load unit if enabl...

Page 12: ...o escape the Calibration menu at any time press ESCAPE The display appears as follows Selecting Cancel will revert back to the Calibration setup Selecting Exit w o saving will return to the menu witho...

Page 13: ...y appears as follows Apply a weight equal to the full scale of the instrument then press ENTER 9 After displaying Please wait the display appears as follows CALIBRATION COMPRESSION Gain adjust Apply f...

Page 14: ...n points have been completed the display appears as follows Press ENTER 12 The display appears as follows Reverse the orientation of the load cell shaft by rotating the gauge 180 degrees Press DIRECTI...

Page 15: ...illating or vibrating in any manner Then try again The calibration weight does not match the set value The entered calibration point is too close to the previous point CALIBRATION TENSION Load too clo...

Page 16: ...m the menu The display appears as follows Selection Description Off Backlight to be off upon powering on the gauge On Backlight to be on upon powering on the gauge Auto Backlight to be on upon powerin...

Page 17: ...The screen appears as follows 11 5 Initial Mode This section is used to configure the initial mode upon powering on the gauge To access this parameter select Initial Mode from the menu The screen appe...

Page 18: ...f up to 24 hours of continuous use Measurement units lbF gF kgF N depending on model Safe overload 200 of full scale display shows OVER at 110 and above Weight gauge only M3 012 M3 100 0 7 lb 0 33 kg...

Page 19: ...olution Load Cell Deflection Model lbF kgF gF N Load Cell Deflection in mm M3 012 0 12 x 0 0001 50 x 0 05 0 5 x 0 0005 0 005 0 13 M3 025 0 25 x 0 0002 100 x 0 1 1 x 0 001 0 010 0 25 M3 05 0 5 x 0 0005...

Page 20: ...Force Gauges User s Guide 19 12 4 Dimensions IN MM Thread Flat M3 012 M3 100 10 32 UNF 5 16 7 94 M3 200 M3 500 5 16 18 UNC 5 16 7 94 www GlobalTestSupply com Find Quality Products Online at sales Glob...

Reviews: