3
4) Enter/Cancel animal weighing function when
“F3 KEY”=5;
UNIT Key
1) When self-checking, press UNIT Key to set unit conversion coefficient.
2) Unit conversion when weighing.
SECTION 3 OPERATIONS
3.1 Power on/off
Press ON/OFF Key to power on the Indicator and the indicator self-checks, then goes into the weighing state.
3.2 Functions and parameters setting
When self-checking, press ZERO Key to enter parameter setting. Then display shows
“F1 CLK”. Press TARE
Key to move functions cycle. Press FUN Key to enter the desire function or MOD Key to cancel or quite. See
details on Section 4 Functions and Parameter Sections.
When self-checking, press TARE Key to enter
“P8 CAL” calibration function.
When self-checking, press MOD Key to enter
“P13 UT” unit setting.
When self-checking, press FUN Key to stop self-checking and it goes into the work state.
When self-checking, press UNIT Key to enter unit conversion coefficient setting.
3.3 Zero the scale
Press ZERO Key to set the zero point from which all other weighing and counting is measured. Usually this will
only be necessary when the platform is empty. When the zero point is obtained the ZERO indicator is ON.
The scale has an automatic re-zeroing function to account for minor drifting or accumulation of material on the
platform. However you may need to press ZERO Key to re-zero the scale if small amounts of weight are shown
when the platform is empty.
3.4 Tare the scale
When a container is applied it needs to tare off and obtains a net weight. Remove all weights from the platform
and make the platform empty. Make sure that the ZERO indicator is on. Place a container on the platform and the
container’s weight will be displayed. Press TARE Key to tare the scale. The container’s weight will be stored as a
tare value and the value will be subtracted from the display, leaving a zero value on the display. The TARE