M-Series Water Purification Machines
1238160 Rev. D
6 - 3
Troubleshooting
TROUBLESHOOTING GUIDE
Symptom
Possible Causes
Remedies
Low operating
pressure
Insufficient feedwater pressure or
flow
Clogged pre-filter
High flow rates
Dirty or fouled sepralators
Solenoid valve not opening
Pump rotating backwards
(three-phase power only)
Insufficient electrical power
Pump not operating correctly
Open the feed pressure,
open the feedwater
valve, check for
restrictions.
Replace the cartridge
Close the concentrate
valve, check the
permeate and
concentrate flow rates
and adjust if necessary.
Excessive permeate flow
may indicate a damaged
O-ring.
Flush and clean the
sepralators.
Clean or replace the
solenoid valve
Switch any two three-
phase leads to the motor
starter.
Check the fuses or
circuit breakers,
measure the voltage.
See the pump
instructions
Summary of Contents for M4-13200
Page 2: ...This page intentionally left blank 1238160 Rev D 23May11...
Page 10: ...M Series Water Purification Machines 1238160 Rev D 1 6 Description Typical P ID...
Page 15: ...M Series Water Purification Machines 1238160 Rev D 1 11 Description NOTES...
Page 31: ...M Series Water Purification Machines 1238160 Rev D 3 9 Preparation and Start Up NOTES...
Page 40: ...M Series Water Purification Machines 1238160 Rev D 4 8 Operation and Maintenance NOTES...
Page 46: ...M Series Water Purification Machines 1238160 Rev D 5 6 Accessories NOTES...
Page 54: ...M Series Water Purification Machines 1238160 Rev D 6 8 Troubleshooting NOTES...
Page 58: ...M Series Water Purification Machines 1238160 Rev D A RGA Procedure NOTES...
Page 59: ......