Table Calibration
This option lets you calibrate direct from the load cell manufacturers certificate without having to apply weights to
the input.
Choose 2 points from the load cell certificate and enter the mV/V values and the weights on the certificate (or
convert the certificate weight to any other engineering unit of your choice) click the ‘
Calibrate’
button and the
calibration is complete.
Engineering Units
This allows you to specify the engineering units that the module was calibrated in. Changing this does not affect the
calibration. It simply assigns the engineering units to the calibrated value so that later the user can select different
engineering units and the module output value will be automatically converted to the selected units.
If you do not want the user changing the units you can hide access to that feature on the
Lock Page
.
If you do not want to display any engineering units then select Undefined from the list and no units will be displayed
next to the value on the Information Page and the user will not be able to select or change the displayed
engineering units.