HCS-3600/3602/3604

Laboratory Grade & High RFI Immunity Switching Mode Power Supply with Rotary Encoder Control

User Manual

1.

     

  INTRODUCTION

 

 

This family of efficient, upgraded SMPS with small form factor, auto cross over CV CC, 3 VI presets, remote control
offers unique solution for various loading conditions and applications.
The dual action (coarse/fine tune) rotary encoder tuning with MCU makes setting the voltage and current levels ever so
smooth, precise and fast. Setting, changing, and checking the current limit level can be done easily without sparking the
output poles.
The remote control functionality allows the output power on/off, voltage &current be adjusted without touching the front
panel of the power supply.

It is suitable for a wide range of applications such as laboratory, telecommunication,  production testing, field test of
voltage critical distant load, powering of dc network and etc...

The 3 user defined presets facilitate quick access to frequently used VI settings.

2.

     

  WARNING

 

 

Do not use this power supply near water.

Do not operate or touch this power supply with wet hands.

Do not open the casing of the power supply when it is connected to ac mains.

Refer all servicing to qualified service personnel only.

Before replacing the AC fuse at AC socket , find out and clear up the cause first.

Replace the AC fuse with the same type and rating as the original fuse.

The max. output voltage of Model HCS-3604 is over 60VDC, avoid touch the metal contact part of the output
terminals.

3.

     

  CAUTION

 

 

Use a grounded 3 pin AC source.

This unit is for indoor use only.

Do not operate or place this unit in a humid, dusty, in direct sunlight location or near any heat source.

Before plugging into local AC mains, check with the rating label at the back of the unit.

•  Do not block any ventilation openings of the unit.

This unit must be used within the specified rating, regular excessive continuous loading may cause damage to the
power supply.

•  Never short the Remote Sensing Terminal.

The gauge size of input power cable must be at least 0.75mm

2

 and the total length of power cable must not exceed

3m.

4.

     

  O

   PERATION ENVIRONMENTAL CONDITION

 

 

10-80%  R.H.

Altitude up to 2000m

Installation category: CAT 2

Pollution degree: 2

Mains supply voltage fluctuation up to ±10% of the normal voltage

Summary of Contents for HCS-3600 USB

Page 1: ...o not operate or touch this power supply with wet hands Do not open the casing of the power supply when it is connected to ac mains Refer all servicing to qualified service personnel only Before replacing the AC fuse at AC socket find out and clear up the cause first Replace the AC fuse with the same type and rating as the original fuse The max output voltage of Model HCS 3604 is over 60VDC avoid ...

Page 2: ...3600 The total rated current Aux Main is 60A HCS 3602 The total rated current Aux Main is 30A HCS 3604 The total rated current Aux Main is 15A 7 Output Terminal Rated 60A for HCS 3600 Rated 30A for HCS 3602 Rated 15A for HCS 3604 8 Remote Sensing Terminal for model HCS 3600 only Warning Shorting the remote sensing terminal or Connect sensing terminal in reverse polarity will damage the power suppl...

Page 3: ... 15V HCS 3602 25V HCS 3604 55V Maximum 6 3 Set Mode First enter into the Set Mode by pushing Switch 9 to Set Mode slot The power supply is then ready to be preset 6 3 1 To define the preset output P1 P2 P3 a Select the Recall Switch 10 to the position P1 P2 or P3 which you want to set b Adjust the front panel voltage control knob to set your desired voltage value c Adjust the front panel current c...

Page 4: ...d make sure it complies with your AC mains voltage Connect the power supply to the AC Mains using the provided power cord Make sure the Mode Switch 9 is at Normal Position Below table to show the self test sequence Self test display and Sequence Test contents To show software version Segment check C V Indicator check C C Indicator check Rear control indicator check Return to C V Start to check Ove...

Page 5: ...rity input of the equipment 7 6 Switch on the power supply first and the panel meter green CV Indicator should light up again 7 7 Switch on the equipment and the panel meter green CV Indicator should still remain in green 7 8 You can now operate the equipment When an operation is finished switch off the equipment first and then switch off the power supply 8 REMOTE SENSING ONLY HCS 3600 HAS THIS FU...

Page 6: ...put voltage range of the power supply by varying the external voltage source Short circuit the main output with 8AWG wire to check the display for CC setting varying the external voltage source 9 2 Method B Using two 0 5K Ohm variable resistors Remark variable resistors 5Kohm Remote Socket Pin Assignment for variable resistor PIN FUNCTIONS REMARKS 1 Internal DC 5V Resistor end 2 Voltage Adjust Var...

Page 7: ...e modes Normal Preset Remote and Set mode A By default Pin 5 is open and output is on B Shorting Pin 5 to Pin 4 ground and output is off C When output is off the LED will show OP OFF Remark using the 8pin remote plug provided and connect with 22AWG wires Pin numbers are marked on the black portion ...

Page 8: ...unit and remove all loading Check your load and output setting Allow the unit to cool down for at least 30 minutes Check if any of the ventilation is blocked check enough clearance around power supply Listen carefully for the short wind noise from the cooling fan when you turn on the unit again If you cannot hear this routine self test wind noise on switch on the fan is fault and do not use the po...

Page 9: ...4 5A Efficiency 85 86 88 Switching Frequency 65 85Khz 75 95Khz 65 85Khz Tracking Over Voltage Protections O P 1 5V set voltage 2V O P 5 15V set voltage 3V O P 1 5V set voltage 2V O P 5 20V set voltage 3V O P 20 30V set voltage 4V O P 1 5V set voltage 2V O P 5 20V set voltage 3V O P 20 60V set voltage 4V Transient Response Time 50 100 Load 1 5ms Power Factor Control Power factor correction 0 95 at ...

Reviews: