2
Oct 02
Fig. 5
•
Position your pillow on the Sit-u-Up as indicated
(fig. 5) and wrap the flap (H) completely around
the pillow to hold it in place.
OPERATING INSTRUCTIONS
•
To raise the Sit-u-Up, press ‘up’
on the
hand control. Release the button to stop at any
point.
•
To lower the Sit-u-Up, press ‘down’
on the
hand control.
FOR YOUR SAFETY:
Ensure that the Airflo compressor, air hose and hand control lead are kept tidy
and do not present a tripping hazard.
ROUTINE CLEANING
Remove the overlay and inflate the Sit-u-Up. Sponge clean with a detergent solution and dry thoroughly.
Do not dry using hot air.
The overlay cover can be machine washed up to 74
°
C or Dry Cleaned.
Do not bleach or iron
.
DISINFECTING
The overlay should be removed and the Sit-u-Up fully inflated. The Sit-u-Up should be pre-cleaned by
washing with detergent and water and then disinfected using either buffered hypochlorite (such as
‘Milton’ solution) or a hard surface alcohol impregnated wipe.
Do not dry using hot air.
STORAGE
Ensure that the Sit-u-Up is thoroughly dry before storing. Store deflated in a dry place. Keep these
instructions in a safe place for future reference.
SPECIFICATIONS
–
Weight (inc. carrying bag): 4.8kg
Max. Temperature: 50
°
C.
Max. Pressure: 0.35 bar (5 psi).
Max. User weight 150kg. (Use with heavier persons will not
damage the product, but full lift may not be achieved).
For further information, please contact your local supplier or the Customer Service Department at
Mangar International, Presteigne, Powys, LD8 2UF, Wales.
Tel: +44(0)1544 267674 Fax: +44(0)1544 260287 E-mail: [email protected]
Website: www.mangar.co.uk
H
70cm
57cm
75cm
65°
(approx)
73cm