All parts which require maintenance work are housed in the gas conditioning system in such a way so
that they are easily accessible. These are (see figure 2):
The filter element of the preliminary filter
FP-2T
.
N O T E !
In order to protect the analysers which have been additionally connected, it
is recommended that in the event of a condensation irruption the moist filter
elements be replaced.
The preliminary filter for the peristaltic pump
PF2
. If the condensate contains particle residue, the
preliminary filter should be replaced at regular intervals. The ‘one-way’ filter is situated in the suction
side of the pump hose (see figure 2) and can be easily replaced;
Hoses of the Condensate pump
SR25.1
. These should be checked every six months and, if
necessary, replaced;
Diaphragm of the sample gas pump
N3KPE
or
N9KPE
. These should be checked every six months
and, if necessary, replaced;