9-3.20-MD
Gas sampling and gas conditioning technology
7
6
TECHNICAL SPECIFICATIONS
PMA05 MB10
PMA05 MB01
Part No.
01A0510
(lowest measuring range
0-10 vol.% O
2
)
01A0520
(lowest measuring range
0-1 vol.% O
2
)
Measuring range
0-100 vol% O
2
(standard), other
measuring ranges according to
client requirements
0-25 vol% O
2
(standard), other
measuring ranges according to
client requirements
Measurement output
0-10V DC non-isolated and 4-20mA for the chosen range, non-isolated,
max. load 300
Ω
Time for 90% value
<3 seconds at 60 l/h
Influence of barometric
pressure
the O
2
display varies proportionally with barometric pressure.
Heating according to each
selected transmitter
55 ° C
Temperature status contact
on transmitters with heating
Contact load max. 48V 1A AC/DC
Contact closes at Temp. > 50 ° C
Accuracy after calibration
deviation: analogue signal output = ±1% of span at range 3-100% / digital
indicator = ±0,1 vol.% O
2
analogue signal output = ±2% of span at range 1%
Sample gas outlet pressure
the sample gas must flow without pressure out to atmosphere.
Influence of sample gas flow
variation in gas flow between 0-60 Nl/hr air will cause a difference of < 0,1
vol.% O
2
Flow rate of sample gas
max. 60Nl/hr
Influence of ambient
temperature
no influence up to +45°C
Ambient/measurement gas
temperature
-10 ° C to +45 ° C
Storage temperature
-20 ° C to +60 ° C.
Ambient humidity
0-95% RH.
Electrical connection
12-pin connector
Power supply
24V DC max. 1.5A
Materials in contact with
medium
Platinum, glass, stainless steel 1.4571, viton, polypropylene, epoxy
Gas connections
Hose connection 4/6, other possible
Dimensions
115 x 115 x 112 mm (HxWxL)
Weight
1kg
Options
PT100 for thermostated version for temperature display or monitoring
Other measuring range than standard range
Zero suppression for spreading of measuring range
For more technical data please refer to the data sheets of your chosen transmitter.