12.1.1
MECHANICAL ZERO POINT ADJUSTMENT
The mechanical zero point is set as follows:
Assign zero gas as described in 13.1.
Loosen fixing screw Item 4 of the photocell bracket.
Turn the adjustment screw Item 6 of the photocell clockwise or counter-clockwise until 0.0%
oxygen is indicated at the signal output;
Retighten fixing item 4 of the photocell bracket;
Illustration 5 Setting the mechanical zero point
Item 4
Item 6