DESCRIPTION
Figure 2
Dimensions and possible construction CSS-VC..
Fine filter
FPF-0, 1GF
Flow meter
FM40
with and without forked light barrier
FA-20mo
Multifunctional expandable controller unit
TCU
Peristaltic pumps
SR25.2
with condensate output at the pump
Sample gas inputs and outputs DN4/6 in connection plate or sample gas inputs on the heat
exchanger
Connection of group alarms (temperature, flow, liquid penetration)
Cold-device socket
3 or 5-way ball valve for calibration gas feeding or
3L/PV
or
5L/PV
M&C
bus interface
Aerosol filter
CLF
..
All components of the gas conditioning systems are installed in a strong compact steel housing for wall
or 19" mount. The ventilation grilles in the side walls provide adequate forced ventilation.
Filter, flow meter and peristaltic pumps are located in the front panel and thus ensure very simple main-
tenance. Due to a removable lid and a hinged front panel, quick and easy inspection and maintenance
of all other built-in components is possible, in particular the easy changing of heat exchangers.
The compressor gas cooler can be equipped with one or two heat exchangers made of glass, stainless
steel or PVDF.
Peristaltic pumps of type
SR25.2
are used for continuous condensate removal or as a metering pump.
The current cooler temperature is displayed on the multi-functional control unit
TCU
mounted on the
front panel. The setpoint temperature of the cooler can be set by pressing keys. In the event of warnings