1. Find the charger port on the
PlasmaFlow device.
2. Plug the charger into the charging
port on the PlasmaFlow device. Plug
the power supply end of the charger
into an electrical wall outlet.
3. The charging LED indicator will
illuminate RED when the battery is
charging. When the batter is fully
charged, the indicator will be
BLUE.
4. If the device is on and
charging, the LED indicator
on the device will be BLUE.
Step 1
Step 2
H
OL
D F
OR
3 SE
CON
DS
Right Leg Pictured
Step 3
Getting Started
Charging PlasmaFlow
PlasmaFlow has a rechargeable lithium-ion battery that provides approximately 7-9 hours of treatment before needing to be
recharged. It takes approximately 3 hours to charge from depleted state. To ensure PlasmaFlow is functioning properly, the
device monitors battery voltage and electrical signal. Important: Charge both device cuffs before first use.
Warning:
Use only the charger provided by ManaMed. The use of the wrong charger can cause excessive heat, damage to
the circuit and shorten the life of the battery.
When Device is OFF:
Plug in the power supply adapter to the wall socket using the plug located on the bottom end of the
device. The RED “Charging” LED indicator (located above the Power Button) on the device will illuminate or flash, depend-
ing on the state of the charge. When the battery is charging, the LED indicator will be RED. Once the battery is fully
charged, the LED indicator will be solid BLUE.
When Device is ON:
The AC Adaptor can be connected while the device is in use. Plug the charger into the bottom of the
controller on the device. Plug the end of the charger into an electrical wall outlet.
Quick Start Instructions for Wearing and Operating PlasmaFlow
Step 1: Calf Cuff Application
Wrap the cuff around each calf
and secure the Velcro to hold it
in place. Make sure the wrap is
snug, but not too tight.
Step 2: Turning the Device On
When the cuffs are secured on
your each of your calves, press
the Power Button for three
seconds until the blue light is
illuminated on each unit. The unit
will be in the first working mode.
Step 3: Starting Treatment
After a couple seconds, pumps inside
the unit will cause the cuff sleeves to
inflate to a pre-set pressure of 55
mmHG. Once the pressure reaches
the proper level, the pumps will turn
OFF for a 50 second ‘rest’ period.
The unit will inflate and deflate to the
specified mode as directed by your
physician. For instructions on how to
switch between modes, refer to the
“Switching Modes” section on page 5.
4