GUIDANCE AND MANUFACTURER’S DECLARATION - ELECTROMAGNETIC EMISSIONS
The PlasmaFlow is intended for use in the electromagnetic environment specified below.
The customer or the user of the PlasmaFlow should assure that it is used in such an environment.
RF Emissions CISPR11
Harmonic Emissions IEC
61000-3-2
Voltage Fluctuations IEC
61000-3-3
RF Emissions CISPR11
Class B
Class A
Complies
Group 1
The PlasmaFlow is suitable for use in all establishments, including domestic establishments and those directly connected
to the public low-voltage power supply network that supplies buildings used for domestic purposes.
The PlasmaFlow uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
Emissions Tests
Compliance
Electromagnetic Environment Guidance
GUIDANCE AND MANUFACTURER’S DECLARATION - ELECTROMAGNETIC IMMUNITY
The PlasmaFlow is intended for use in the electromagnetic environment specified below.
The customer or the user of the PlasmaFlow should assure that it is used in such an environment.
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Electrical Fast
Transient/Burst
IEC61000-4-4
Surge
IEC61000-4-5
±1kV differential
mode
±2kV common mode
±1kV differential
mode
±2kV common mode
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment. If the user of the
PlasmaFlow requires continued operation during power mains interruptions, it is recommended that the
PlasmaFlow be powered from an uninterrupted power supply or a battery.
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC61000-4-11
Power Frequency
(50/60Hz)
Magnetic Fields
IEC61000-4-8
30 A/m at
50 or 60 Hz
30 A/m at
50 or 60 Hz
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical
commercial or hospital environment.
NOTE: UT is the a.c mains voltage prior to application of the test level.
<5%UT (>95% dip in UT)
for 0.5 cycle
40%UT (60% dip in UT)
for 5 cycles
70%UT (30% dip in UT)
for 25 cycles
<5%UT (>95% dip in UT)
for 5 seconds
<5%UT (>95% dip in UT)
for 0.5 cycle
40%UT (60% dip in UT)
for 5 cycles
70%UT (30% dip in UT)
for 25 cycles
<5%UT (>95% dip in UT)
for 5 seconds
±2kV for power
supply lines
±1kV for input/
output lines
±2kV for power
supply lines
±1kV for input/
output lines
Mains power quality should be that of a typical commercial or hospital environment.
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material,
the relative humidity should be at least 30%.
±8kV contact
±15kV air
±8kV contact
±15kV air
Immunity
Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic Environment Guidance
Compliance Statements & Declarations Electromagnetic Compatibility (EMC) Tables