background image

Model PR-264

Technical Information

TI.264-04

PRESSURE TRANSDUCER

Page 4 of 4

CHECKOUT

DIMENSIONAL DATA

CALIBRATION

1.  Verify that the unit is mounted in the correct position.

2.  Verify appropriate input signal and supply voltage.

3.  Verify appropriate configuration range.

CAUTION:  Never connect 120 VAC to these transducers. 
Never connect AC voltage to a unit intended for DC supply.

Transducer
Operation

This is a rough functional check only.

1.  Adjust the pressure to obtain maximum output signal for 

appropriate range.

2.  Output should be 20 mA or 5 or 10 VDC.

3.  Adjust the pressure to obtain minimum output signal.

4.  Output should be 4 mA or 0 VDC.

Calibration of PR-264 mA Units

1.  Connect terminals [+] and [-] to the appropriate power 

source.

2.  Connect the DVM in series on the [-] terminal.

3.  Apply low pressure to the unit and carefully adjust the zero 

trimmer [Z] to obtain desired low output.

4.  Apply high pressure to the unit and adjust span trimmer [S] 

to obtain the desired high output pressure.

5.  Repeat steps 3 and 4 until desired calibration is achieved.

All units are factory calibrated to meet or exceed published 
specifications. If field adjustment is necessary, follow the 
instructions below.

MAINTENANCE

Regular maintenance of the total system is recommended to 
assure sustained optimum performance.

FIELD REPAIR

None. Replace with a functional unit.

WARRANTY

See Data Sheet for additional information.

Calibration of PR-264 VDC Units

1.  Connect terminals [+] and [-] to the appropriate power 

source. The [-] terminal is also the negative output terminal.

2.  Connect the DVM  on DC volts across [0] and [-] terminal.

3.  Apply low pressure to the unit and carefully adjust the zero 

trimmer [Z] to obtain desired low output.

4.  Apply high pressure to the unit and adjust span trimmer [S] 

to obtain the desired high output pressure.

5.  Repeat steps 3 and 4 until desired calibration is achieved.

Figure 7 – PR-264 Pressure Transducer dimensions shown in inches and 
millimeters (mm).

For Technical / Application Assistance call your nearest office

8189 Century Boulevard • Minneapolis, MN 55317-8002 • USA

800-843-5116 • 952-556-4900 • Fax 952-556-4997

[email protected] • www.mamacsys.com

EUROPE

Baird House, Units 6 & 7

Dudley Innovation Centre

Pensnett Estate • Kingswinford

West Midlands • DY6 8XZ

United Kingdom

01384-271113 • Fax 01384-271114

AUSTRALIA

4 Arminger Court, Unit 2

Holden Hill • S.A. 5088

Australia

08-8359-4333 • Fax 08-8395-4433

CANADA

155 McIntosh Drive, Unit 5

Markham • Ontario • L3R 0N6

Canada

905-474-9215 • Fax 905-474-0876

ASIA

No. 22 Lorong 21A Geylang #11-02

Prosper Industrial Building

Singapore • 388421

656-3927273 • Fax 656-3927276

MAMAC Systems, Inc., reserves the right to change any specifications without 

notice to improve performance, reliability, or function of our products.

NOTE:  The PR-264 is a highly accurate device.  For 
applications requiring a high degree of accuracy, the use of
laboratory quality meters and gauges are recommended.

TI.264-04

RoHS

RoHS

RoHS

Summary of Contents for PR-264

Page 1: ... Disconnect power supply before installation to prevent electrical shock and equipment damage Make all connections in accordance with the job wiring diagram and in accordance with national and local electrical codes Use copper conductors only Caution Use electrostatic discharge precautions e g use of wrist straps during installation and wiring to prevent equipment damage Avoid locations where seve...

Page 2: ...ers with mA output 1 Remove the terminal block by carefully pulling it off the circuit board 2 Locate the and terminal markings on the board 3 Attach the supply voltage to the lead 4 Connect the 4 20 mA output terminal to the controller s input terminal 5 Ensure that the power supply common is attached to the common bus of the controller 6 Re insert the terminal block to the circuit board and appl...

Page 3: ... 2 O N O N 1 2 O N O N 1 2 O N O N 1 2 O N O N 0 500 psig 0 3500 kPa default 0 250 psig 0 1750 kPa 0 125 psig 0 875 kPa 1 2 O N O N 1 2 O N O N 1 2 O N O N 0 100 psig 0 700 kPa default 0 300 psig 0 2000 kPa default 0 50 psig 0 350 kPa 0 25 psig 0 175 kPa 0 150 psig 0 1000 kPa 0 75 psig 0 500 kPa 1 2 O N O N 1 2 O N O N 1 2 O N O N 1 2 O N O N 1 2 O N O N 1 2 O N O N 0 500 psig 0 3500 kPa default 0...

Page 4: ... Data Sheet for additional information Calibration of PR 264 VDC Units 1 Connect terminals and to the appropriate power source The terminal is also the negative output terminal 2 Connect the DVM on DC volts across 0 and terminal 3 Apply low pressure to the unit and carefully adjust the zero trimmer Z to obtain desired low output 4 Apply high pressure to the unit and adjust span trimmer S to obtain...

Reviews: