Chapter 2
Site requirements
Page 2-4
MAN 0382
Power requirements
The mains power supply must be clean and filtered. If necessary, fit an Un-
interruptible Power Supply (UPS) to remove any spikes or noise. The power
requirements are:
Power cords and Power safety
The notes in this section indicate best practice. Adhere to these when connecting
the instrument to the power supply.
Warning!
Do not operate this product with a damaged power cord set. If the power
cord set is damaged in any way, replace it immediately.
Warning!
Do not use the power cord supplied with this product on any other
products.
Power cord set requirements
Power cord sets must meet the requirements of the country where the product is
used. For more information, contact a Malvern representative.
Component
Width
Depth
Height
Zetasizer Nano
320mm
600mm
260mm
MPT-2 Autotitrator
170mm
390mm
260mm
Computer and printer
See manufacturer’s documentation
Component
Power requirement
Power sockets required
Zetasizer Nano
~ 100-240V, 50-60Hz
1
MPT-2 Autotitrator
~ 100-240V, 50-60Hz
1
Vacuum Degasser
~ 100-240V, 50-60Hz
1
Computer
See manufacturer’s documentation
Computer monitor
See manufacturer’s documentation
Printer
See manufacturer’s documentation
Man0382-5.0 (Zetasizer Nano Essentials).book Page 4 Wednesday, December 2, 2009 12:06 PM