Site requirements
Chapter 2
Zetasizer Nano Essentials
Page 2-3
Introduction
This document outlines site requirements for a Zetasizer Nano. Ensure all these
are met
before
the instrument is delivered.
Environmental conditions
The site must be:
Indoors and away from strong light (windows).
Away from heat sources like radiators.
Well ventilated (for noxious materials).
On a horizontal vibration-free bench.
We recommend that the computer is positioned to the right of the instrument.
Space required
Provide enough space to allow easy access to all components and connections. This
diagram shows the recommended space for a system. Always place the Autotitrator
to the right of the instrument.
ill 7931
Allow at least
800mm
above the bench surface for access to the cell area. The
system dimensions are given below (the width is with the cuvette holder closed):
300 mm
150 mm
150 mm
300 mm
1240 mm
750 mm
800 mm
(maximum)