10
2. TPress
①
(ON/OFF) button.
(Main page)
3. Select test voltage:
3.1. Select one from 1kV, 5kV, 10kV or 15kV, press (
②
,
③
,
④
or
⑤
) respectiv e l y .
3.2. To select voltage other than the four indicated, press
anyone among
②
,
③
,
④
,
⑤
, then, press
⑥
(volt-add) or
⑦
(volt-reduce) till the require
d
voltage is reached.
4. Be sure that the cable connecting the test object and
Tester is correct. Then, press
⑧
(TEST/STOP); LCD
displays as below:
5. Then, press
⑧
(TEST/STOP).
Note:
a. During test process, if there is a exterior voltage (above
AC30V or DC30V) exerted, beeper activates in response;
and LCD displays the warning picture as shown below:
Now, the test cannot be performed. To go on with the test
remove the exterior voltage.
b. While test is running, beeper activates to remind the
operator that test is underway.
c. After the set test time is due (see Function 3: the test time
setting), test stops and system will automatically lock
down the test value.
d. To read the test value on the LCD display, press the
⑪
BACKLIT button( ).