Maico ERO SCAN Operating Instructions Manual Download Page 29

 

Operating Instruction  
ERO•SCAN

®

 

Screening and Diagnostic Version

 

 

operating_instructions_8106556-6_EROSCAN_e_15a 

 

22 

8106556-5   02/15 

 

 

10.8

 

Viewing Results 

 

When  testing  is  complete,  a  Display similar  to  Display 8  will 
appear.  The  results  of  the  test  are  automatically  saved  in 
memory  as  soon  as  the  test  is  complete.  The  results  will  be 
saved even if the unit turns off or the batteries are temporarily 
depleted.  This  screen  again  indicates  the  test  ear  and  further 
gives the results of the test.  
 

 

“PASS” on the screen indicates  that the patient passed the 
screening.  

 

“REFER”  indicates  that  the  patient  did  not  pass  the 
screening.  

 

“NOISY”  indicates  that  excessive  noise  was  present  during 
the test.  

 

“NO  SEAL”  indicates  that  a  seal  was  not  maintained 
throughout the test.  

 

“FIT  ERR”  indicates  inadequate probe  placement in  the  ear 
canal to produce target stimulus intensities.  

When the test result is “NOISY,” “NO SEAL,” or “FIT ERR” the tester should reposition the 
probe, selecting a different size eartip if necessary, and retest. If the test result is “REFER” 
the tester can determine from the printout whether the test should be repeated. To review 
the results, push the 

DOWN arrow key to return to the graph. After reviewing the results, 

again push the 

DOWN arrow key to return to the Main Menu (Display 1, page 19).  

 

NOTE:  Completed  tests  are  automatically  saved.  By  default  (Save  L/R  Mode),  the 
ERO•SCAN

®

 instrument will save only the last test for each ear. Starting a new test for the 

same ear will overwrite the existing test. 

 

NOTE:  Once  the  review  screen  is  exited,  results  are  no  longer  viewable  within  the  device.  
Printing or transferring to the PC is required for continued viewing options. 

 

See  the  next  section 

Managing  Results  on  page 24  for  more  information  on  how  the 

ERO•SCAN

®

 saves results. 

 
See the section 

Instrument Settings – Save Mode on page 40 for more information on the 

ERO•SCAN

®

 save mode options. 

10.9

 

Test Technique 

 
As  with  other  otoacoustic  emission  test  instruments,  there  is  a  technique  to  learn  when 
using the ERO•SCAN

®

 instrument, especially while testing infants. Experience with existing 

OAE systems suggests that it may take up to 3 months to become completely proficient at 
screening infants.  
 

 

Display 8 

 

Summary of Contents for ERO SCAN

Page 1: ...MAICO Diagnostic GmbH Salzufer 13 14 10587 Berlin Germany Tel 49 30 70 71 46 50 Fax 49 30 70 71 46 99 Operating Instruction ERO SCAN Screening and Diagnostic Version...

Page 2: ......

Page 3: ...8 5 3 Warning In Case of Emergency 8 5 4 Warning Explosion 8 5 5 Warning Battery Safety 8 5 6 Warning General 8 6 Recycling Disposal 9 7 Regulatory Symbols 10 8 Introduction 12 8 1 How Does the ERO S...

Page 4: ...mal Printer 25 11 4 Connecting to the PC Database Software 26 12 MPT II Mobile thermal mini printer 27 12 1 Operation 28 13 Interpreting Results 32 13 1 Understanding the Display 32 13 2 SNR Graph Vie...

Page 5: ...Protocol 45 18 7 Save Protocol 45 19 Advanced Options for TEOAE Testing 45 19 1 Instructions for Customizing a Test Protocol 46 19 2 Selecting the Averaging Time 47 19 3 Setting the PASS SNR Level 47...

Page 6: ...Electrical Safety EMC and Associated Standards 64 Guidance and Manufacturer s Declaration Electromagnetic Emissions 65 Recommended Separation Distances between Portable and Mobile RF Communications Eq...

Page 7: ...et www maico diagnostic biz EC REP Copyright 2014 MAICO Diagnostic All rights reserved No part of this publication may be reproduced or transmitted in any form or by any means without the prior writte...

Page 8: ...5 3 2007 Electroacoustics Audiometric equipment Part 3 Test signals of short duration IEC 60645 6 2010 Electroacoustics Audiometric Equipment Part 6 Instruments for the measurement of otoacoustic emis...

Page 9: ...OTHER WARRANTIES EXPRESSED OR IMPLIED INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE For assistance with this ERO SCAN Otoacoustic Emission Te...

Page 10: ...sult when a three prong to two prong adaptor is connected between the ERO SCAN power supply and an AC outlet No modifications of the equipment are allowed by anyone other than a qualified MAICO repres...

Page 11: ...mpatibility and are listed as meeting the requirements of the UL Medical and Dental Equipment Standard Connection of accessories not meeting these requirements may result in electrical leakage current...

Page 12: ...tests are within 5 tests of the maximum limit Memory Full The maximum saved test limit is reached The user will need to clear the memory before any additional tests can be performed Power Low The bat...

Page 13: ...system components or accessories the unit should not be used until deemed safe by a MAICO certified service technician Do NOT use in the presence of flammable gaseous mixtures Users should consider th...

Page 14: ...to be used by hearing healthcare professionals i e ENT doctors audiologists and or technicians neonatal nurses and school nurses who have been trained by a hearing healthcare professional 5 1 Caution...

Page 15: ...Warning In Case of Emergency In case of emergency disconnect the instrument from the supply mains by removing the micro USB cable from the connector as shown in Figure 2 on page 16 5 4 Warning Explos...

Page 16: ...ements of electronic devices Follow all your respective local laws and regulations for the proper disposal of batteries and any other parts of this system Below is the contact address for proper retur...

Page 17: ...nufacturer 980 60601 1 Caution Consult Accompanying Documents 980 60601 1 Return to Authorized Representative Special Disposal Required 980 60601 1 Reference Number 60601 1 B Patient Applied Part Acco...

Page 18: ...Operating Instruction ERO SCAN Screening and Diagnostic Version operating_instructions_8106556 6_EROSCAN_e_15a 11 8106556 5 02 15...

Page 19: ...they are printed The plastic housing contains circuit boards that provide the signal processing and Display the test results The instrument also contains a rechargeable lithium ion battery to power t...

Page 20: ...latile memory so the operator can delay printing until a later time if desired 8 3 Sensitivity and Specificity Sensitivity and specificity of this type of device are based on the test characteristics...

Page 21: ...tional accessories or Sanibel disposables contact your special equipment distributor or MAICO Diagnostic at 49 30 70 71 46 50 We recommend that you save the shipping box and packing materials in case...

Page 22: ...Operating Instruction ERO SCAN Screening and Diagnostic Version operating_instructions_8106556 6_EROSCAN_e_15a 15 8106556 5 02 15...

Page 23: ...of the charging function and is defined as follows Steady state illumination indicates that the battery is fully charged This identifies that the charging cycle has been completed or was not implement...

Page 24: ...artips are disposable and should be replaced after each patient After selecting an eartip push it onto the probe tube until it is flush against the base of the probe tube Figure 5 Twisting the eartip...

Page 25: ...ndicating the instrument is ready for use The Flash Screen will appear briefly This Display will indicate the type of instrument Screener SCR Screener Plus SC Standard STD or Combo CMB software versio...

Page 26: ...e selected protocol Press the UP arrow to return to the Main Menu to begin testing Press the SETUP to enter the setup menus For either DPOAEs or TEOAEs screener devices there are two default test prot...

Page 27: ...icates that the probe is not inserted into the ear or placed properly When the bars are all the way to the right Display 4 this represents the volume is too small for the test to begin or indicates th...

Page 28: ...calibration of the test tones the test phase will begin automatically During the test phase a set of bars should appear on the Display Display 6 These are the test results which are displayed as the e...

Page 29: ...nd retest If the test result is REFER the tester can determine from the printout whether the test should be repeated To review the results push the DOWN arrow key to return to the graph After reviewin...

Page 30: ...level in the ear canal is low and it will appear more often if the noise level in the ear canal is high Otoacoustic emissions are very low level sounds Any noise in the ear canal at the time of testin...

Page 31: ...h patient in the default mode When operating in the Save 250 mode the ERO SCAN will save up to 250 tests There are two options in the Save 250 mode 1 The ERO SCAN will automatically number each test f...

Page 32: ...ng is provided that data will be deleted Data can be deleted using the Clear function in the System Menu page 39 Data can be deleted in the ERO SCAN from the Database Software when the device is conne...

Page 33: ...the paper runs out before printing is complete NOTE See the ERO SCAN Database Software Manual for instructions on printing from the Database Software 11 4 Connecting to the PC Database Software Connec...

Page 34: ...on induction cookware In the event that the battery leaks and the fluid gets into one s eye do not rub the eye Rinse well with water and immediately seek medical care If left untreated the battery flu...

Page 35: ...per roll as shown and close the lid Figure 9 NOTE Reorder paper from Maico or your local distributor Battery pack insertion Figure 10 Insert battery as shown Figure 10 Power on Push POWER BUTTON for t...

Page 36: ...ode Self test While printer is powered OFF press and hold PAPER FEED button then press and hold POWER BUTTON simultaneously When beep is heard after approx 3 seconds release both buttons and a test pa...

Page 37: ...ral printers powered on and within range while searching It is possible to pair up to 8 instruments with one printer If more instruments are paired the oldest will be deleted For detailed instructions...

Page 38: ...m 2 2 in 0 04 in max 40 mm 1 6 in diameter Battery pack 2 cell Li Ion battery pack 7 4 V 1500 mAh Power supply charger 12 V 1 A UE15WCP1 120125SPA Maximum current consumption 0 5 A See also Figure 12...

Page 39: ...s complete The display might differ depending on the settings used for the test Figure 133 The display above shows a 6 column graph Figure 13 Each frequency DPOAEs or frequency band TEOAEs is indicate...

Page 40: ...represent a pass a yellow bar is a non Pass at the frequency band NOTE Diagnostic protocols can display green or purple bars Green is displayed when a PASS criterion is set in the device For more info...

Page 41: ...re 14 1 The software version number e g V100 01 2 The time and date of the test based on the setting of the internal clock if the clock is set correctly this time and date will be correct 3 The test n...

Page 42: ...and date will be correct 3 The test number if operating in Save 250 mode e g 001 4 The protocol selected e g TE 32s 5 The averaging time used for this test e g 10 sec avg 6 Instrument Probe serial nu...

Page 43: ...in SNR 4 9 dB The printout values are rounded up to the nearest integer and are shown as TE 5 NF 0 and SNR 5 This can result in what appears to be an error with regard to the pass criterion If the pas...

Page 44: ...ess SETUP at the protocol selection Display Display 2 page 19 The current date and time presently set in the unit will be shown Display 13 If the time and date are correct press the UP key to escape b...

Page 45: ...The ERO SCAN unit can be paired with only one device at any time To establish Bluetooth pairing turn on the device that will be paired with the ERO SCAN unit e g thermal printer Then select DISCOVER t...

Page 46: ...ing Test Results The Test Results Clear menu Display 19 allows the user to clear the test results stored in the unit without printing them Select the LEFT or RIGHT arrow key to clear the results and s...

Page 47: ...starting with test number 1 When patient names are used patient names are uploaded from the ERO SCAN Database Software to the ERO SCAN unit the patient names are displayed on the ERO SCAN Unit in the...

Page 48: ...st Pass To change the mode to Minimum Amplitude setting press the LEFT or RIGHT arrow keys to make a selection Once you have made your selection press NEXT The minimum DP amplitude when ON is 5 dB SPL...

Page 49: ...is shown Press NEXT to exit this menu 17 9 Reset to Default The Reset to Default menu Display 27 will return all instrument settings and protocol settings to their original factory defaults Select the...

Page 50: ...memory so the settings will be retained even when the battery is discharged temporarily 18 1 Instructions for Customizing a Test Protocol To enter the DPOAE Menu 1 Press CHANGE at the main menu Displa...

Page 51: ...s 0 5 sec 1 0 sec 2 0 sec or 4 0 sec Press the CHANGE keys to select an option and the NEXT key to exit Display 29 NOTE Longer averaging times help to reduce the noise floor which can improve the like...

Page 52: ...only 4 frequencies are being tested would result in every test being labeled as a REFER NOTE To disable the PASS REFER indication set the number of frequencies for pass to 0 NOTE Diagnostic protocols...

Page 53: ...so the settings will be retained even when changing the batteries 19 1 Instructions for Customizing a Test Protocol To enter the TEOAE Menu 1 Press CHANGE at the main menu Display 1 page 19 2 Using t...

Page 54: ...onds The instrument will automatically stop the test when the pass criterion is met prior to the averaging time Press the CHANGE keys to select an option and the NEXT key to exit Display 34 19 3 Setti...

Page 55: ...to indicate a PASS Press the CHANGE keys to select an option and the NEXT key to exit Display 36 NOTE Diagnostic protocols display the SNR graph with purple bars Once the Number of Frequencies for PAS...

Page 56: ...is appropriate for this type of instrument This may not conform to the infection control guidelines of the user s facility The disinfection materials and procedures applied in the users facility may b...

Page 57: ...IEC and UL 60601 1 20 3 Probe Tube Replacement Probe tubes are disposable and should be replaced when they become clogged A package of replacement probe tubes is included with this instrument Do not...

Page 58: ...t in your own ear with the proper eartip for testing yourself If the test will not start or if the AutoStart tones sound unusual replace the probe tube Contact MAICO for service if the problem persist...

Page 59: ...blish Bluetooth pairing Device Not Responding The printer is not responding to queries from the instrument Check printer status Awaken printer from sleep mode Charge printer battery if necessary Due f...

Page 60: ...ruption in testing Memory Full The maximum saved test limit is reached The user will need to clear the memory before any additional tests can be performed Power Low The battery charge level is too low...

Page 61: ...SPL at 1 kHz 1 Hz bandwidth Dimensions and Weight Length 1 0 meter 40 in Weight 28 g 1 00 oz Connector HDMI INSTRUMENT SPECIFICATIONS Power Supply Lithium Ion rechargeable Battery Life 1000 tests per...

Page 62: ...use only approved power supply Model No UE08WCP 050160SPA Output 5 0 V DC 1 6 A Input 100 V 240 V AC 50 Hz 60 Hz 400 mA ENVIRONMENTAL REQUIREMENTS Operating Temperature 15 C to 35 C 59 F to 95 F Oper...

Page 63: ...Operating Instruction ERO SCAN Screening and Diagnostic Version operating_instructions_8106556 6_EROSCAN_e_15a 56 8106556 5 02 15 Appendix A Flowcharts A 1 Test Operation Flowchart...

Page 64: ...on ERO SCAN Screening and Diagnostic Version operating_instructions_8106556 6_EROSCAN_e_15a 57 8106556 5 02 15 A 2 Setup Menu Flowchart NOTE DPOAE Menu and TEOAE Menu are not accessible in the ERO SCA...

Page 65: ...main estimates of the actual L1 L2 distortion DP and noise floor NF are obtained via the discrete Fourier Transform with a bin resolution of approximately 31 Hz The NF estimate is obtained by averagin...

Page 66: ...lly when no emission is actually present If a test is performed with the instrument s probe placed in a test cavity it can be shown theoretically that the SNR will be greater than 6 dB approximately 7...

Page 67: ...ncies drops the probability of passing an ear with significant hearing loss to 1 or less NOTE By the binomial distribution two of three frequencies at 8 4 dB or three of six frequencies at 7 dB should...

Page 68: ...ed for DPOAEs Without that difference the numerical SNR value for a PASS with the two methods would be quite similar The ERO SCAN uses a novel noise rejection algorithm patent pending that permits acc...

Page 69: ...62 8106556 5 02 15 The artifact rejection can only reject the noisiest samples in a measurement period If the ambient noise level rises too high and or the eartip seal is poor then all samples will be...

Page 70: ...5 6 6 3 7 1 8 65 55 4 sec 6 dB 0 DP 1 5 12 12 1 5 2 3 4 5 6 7 8 9 10 11 12 65 55 4 sec 6 dB 0 Diagnostic version also includes DP 4s screening protocol TEOAE Protocols Name of Freq Bands Freq center b...

Page 71: ...ng sure the system complies with the IEC 60601 1 2 standard Electromagnetic Compatibility Although the instrument fulfills the relevant EMC requirements precautions should be taken to avoid unnecessar...

Page 72: ...is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the ERO SCAN can help prevent electromagnetic interferences by maint...

Page 73: ...hospital or residential environment Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial hospi...

Page 74: ...ended separation distance in meters m Field Strengthens from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b...

Page 75: ...perating_instructions_8106556 6_EROSCAN_e_15a 68 8106556 5 02 15 Specifications are subject to change without notice MAICO Diagnostic GmbH Salzufer 13 14 10587 Berlin Germany Tel 49 30 70 71 46 50 Fax...

Reviews: