Operating Instruction
ERO•SCAN
®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
40
8106556-5 02/15
seconds, 1, 2, or 4 minutes. Once you have made your
selection, press
NEXT.
17.4
Save Mode/Storing Test Results
The ERO•SCAN
®
unit automatically stores only the most recent
test result for each ear L/R (Display 21), but has the capacity to
store 250 individual tests. To change the mode to save up to
250 tests, press the
LEFT or RIGHT
arrow keys to change the
menu to 250. Once you have made your selection, press
NEXT.
There are two options in the Save 250 mode (Display 22):
1. The ERO•SCAN
®
will automatically number each test
from 1 to 250
2. The ERO•SCAN
®
Database Software is used to transfer
patient names to the ERO•SCAN
®
and the ERO•SCAN
®
will display the names. Up to 50 names can be stored in
the device and 250 tests.
When numbers are used (no patient names are uploaded from
the ERO•SCAN
®
Database Software to the ERO•SCAN
®
), each
test is automatically incremented, starting with test number 1.
When patient names are used (patient names are uploaded from the ERO•SCAN
®
Database Software to the ERO•SCAN
®
unit) the patient names are displayed on the
ERO•SCAN
®
Unit in the same order as displayed on the Database Software. To move to a
different name than the one displayed on the ERO•SCAN
®
screen, use the left or right
arrows to cycle through the names until the desired name is on the display. A patient
named “Unnamed” is always included at the beginning of the ERO•SCAN
®
list for
instances when a patient is being tested, but the patient name was not transferred to the
ERO•SCAN
®
.
It is recommended that you go to the Test Results Clear screen (Display 19, page 39) to
clear any previous memory locations after you changed the save mode setting.
NOTE: When using the “250 test” mode, it is important to keep a record of the test
number for each patient. When 245 tests have been saved, the user will be warned that
the memory is almost full. When the ERO•SCAN
®
unit reaches 250 saved tests, it will not
allow any further testing. At this point either the results must be printed, transferred to the
PC software, or they must be cleared from memory.