Operation Manual easyTone
8529689 Rev. 2
8
10/05/2022
2.5 General Precautions
Before starting a measurement make sure, that the device
works properly.
Use and store the device indoors only. For operation, storage
and transport conditions see Table in Section 6 Technical Data.
For operation in certain places, a recalibration may be
necessary.
No modification of this equipment is allowed.
No part of the equipment can be serviced or maintained while
in use with the patient.
Do not drop or otherwise cause undue impact to this device.
If the device is dropped or otherwise damaged, return it to the
manufacturer for repair and/or calibration. Do not use the
device if any damage is suspected.
Uncalibrated devices may lead to faulty measurements and
may cause the examinee to be exposed to loud sounds.
Ensure that the device does not come into contact with
liquids. Should the user suspect fluids have contacted the
system components or accessories, the unit should not be
used until deemed safe by a MAICO certified service
technician.
2.6 Electrical and Electrostatic Safety
This icon indicates that patient applied parts of the device
conform to IEC 60601-1 Type B requirements.
In Case of Emergency
In case of emergency, disconnect the audiometer
headphone from the tablet.
In case of emergency, disconnect the tablet from the
computer.
In Case of Emergency
In case of emergency, disconnect the device from power
supply.
Position the device in such a way that it can be easily
disconnected from the power supply at any time.
Do not use the device if the power supply unit and/or the
plug is damaged.
WARNING
WARNING
WARNING
WARNING
WARNING
WARNING