Displays Indicating Action Required
This will display if the tank which holds used
fluid fills to capacity.
Refer to instructions for emptying the Used Fluid
Tank in the next section.
This will display if the level of fluid is too low in
the bottle of new fluid.
A new bottle must be selected and installed on
unit.
Empty Used Fluid Tank
1. Disconnect the fluid line coupler and the electrical float switch cable on the Used Fluid Tank.
2. Remove cap on Used Fluid Tank (turn counterclockwise).
3. Empty used fluid into an appropriate bulk waste container.
4. Replace cap on Used Fluid Tank (turn clockwise).
5. Connect fluid line coupler and the electrical float switch cable.
Follow all federal and local laws and regulations when disposing of hazardous material.
Filter Maintenance
The BFX-2 has six filters to prevent debris from entering internal
components.
New Fluid Filter: Filters fluid from the New Fluid Bottle. If the fluid
flow seems slow while trying to fill the master
cylinder - check and clean this filter.
This filter can be accessed through the front
door on the top of the BFX-2. Remove screws
and gently raise the door to access the filter.
Unscrew the filter cover and clean the screen
as required.
Vacuum Filter: Filters fluid in the five vacuum hoses. A filter is
mounted in-line on each of the hoses. If the fluid
flow seems slow from the bleeders - check and
clean these filters.
These filters can be accessed through the rear
door on the top of the BFX-2. Remove the two
thumb screws and lift the door. The filters will be
seen about 12 inches below where the hoses
go through the panel inside. Remove hoses
from both ends of the filter and try clearing the
filter with compressed air.
U S E D T A N K F U L L
M U S T B E E M P T I E D
N E W T A N K E M P T Y
P R E S S C O N T I N U E