Starting a New Measurement
A new measurement is started when
•
the measuring unit is switched on.
•
the force sensor is connected to the measuring unit.
•
the maximum force exceeds 999 N.
•
either button is pushed.
Offset Adjustment
An offset adjustment is carried out when
•
the measuring unit is switched on.
•
the force sensor is connected to the measuring unit.
•
either button is pushed for at least 2 seconds.
Pos: 35 / Technische Dokumentation/Alle Geräte/Überschriften/ Überschriften 1.1/A/Überschrift 1.1: Ab gleich @ 34\mod_ 1382688332978_ 75.docx @ 1846087 @ 2 @ 1
abgleich
2.9
Adjustment
Pos: 36 / Technische Dokumentation/Leistungsmess technik/Z7 Z ubehör allgemein/ 2201 S KM 2/ BA/Inhalt: Z722 Abgleich @ 34\mod_ 1382598744362_75.d ocx @ 1845269 @ @ 1
If the meter is not exactly adjusted, a correction factor must be entered in
variable 5.
1 Load the meter with a known weight (e.g. 30 kg).
The meter display shows the weight in Newton, e.g. 305 N.
In fact, the meter should display 30 kg • 9.81
𝑚
𝑠
2
≈
294 N.
2 Calculate correction factor:
Correction factor =
𝑎𝑎𝑎𝑎𝑎𝑎
𝑣𝑎𝑎𝑎𝑣
𝑚𝑣𝑎𝑠𝑎𝑚𝑣𝑚
𝑣𝑎𝑎𝑎𝑣
• 1000
In this example: Correction factor =
294
𝑁
305
𝑁
• 1000
≈
964
By entering the correction factor in variable 5, the display error can be
compensated for.
3 Open the service menu and select
menu item "EE-User variables".
4 Press the right-hand key to open the
user variable program.
Menu Selection
EE-User Variables
5 Select user variable 5 with the left-
hand key (previous variable) or right-
hand key (next variable).
6 Open the variable with the right-hand
key (long press).
EEPROM Variables
Factor Force
Var
5:
1000