![Magtrol ED-715 User Manual Download Page 34](http://html1.mh-extra.com/html/magtrol/ed-715/ed-715_user-manual_694852034.webp)
31
Magtrol Hysteresis Dynamometers
Chapter 3 – Installation/Configuration
SETUP
3.3
SYSTEM CONFIGURATIONS
After the load cell shipping/restraining bolt has been removed and the dynamometer has been
earth grounded, the unit is ready for connection to the appropriate readout instrument and power
supply.
3.3.1
M
anual
T
eST
S
ySTeMS
The dynamometer can be set up as a manual test system for quick pass/fail testing on the production
line or at incoming inspection. A typical manual test system will consist of a Magtrol Hysteresis
Dynamometer used in conjunction with a Magtrol 6200 Open-Loop Controller.
Note:
Magtrol’s Single or Three-Phase Power Analyzer may be included
as an option.
The following diagram illustrates the required cable sets and connections for a manual test system
setup.
1
o----------
o----
o----------
2 Pin Dynamometer Brake Cable
PN88M070-0150 (included with dynamometer)
6200 CONTROLLER
14 Pin/14 Pin Instrument Cable
PN88M007-0150 (included with dynamometer)
HD / ED
DYNAMOMETER
Figure 3–6 Dynamometer with 6200 Controller