1
・
Functions added by V1.05.00JP
LT80-NE software version V1.05.00JP adds and changes the following functions compared to
V1.04.00JP.
1. Display resolution setting
・・・・・・・・・・・・・・・・・・・・・
P.2
・
The measurement value display resolution can be set in the view frame.
Measurement values can be displayed at a resolution lower than the input resolution
of the measuring unit.
Available display resolution settings: 0.1 µm, 0.5 µm, 1.0 µm, 2.0 µm, 5.0 µm, 10.0 µm
2. Scaling
・・・・・・・・・・・・・・・・・・・・・・・・・・・・・
P.3
・
A function that sets a magnification for measurement display values has been added.
This function sets the magnification for each view frame, and also supports frames
with a calculation set.
Display value = Measurement display value × Set magnification
3. Preset settings for frames with a calculation set
・・・・・・・・・・・・
P.4
・
Preset values can be set for frames with a calculation set.
4. Measurement data save
・・・・・・・・・・・・・・・・・・・・・
P.5
・
Manual operation, display unit I/O operation, or auto save by number of data can be
selected when saving measurement data to a USB memory or SD card.
5. Function change when using the MG80-LM
・・・・・・・・・・・・・
P.10
・
The function has been changed so that normal measurement can be performed while
latch operation is stopped (standby status for operation).
* Functions 1 to 3 above support MG80-MA with serial numbers 300001 and higher.
When using functions 1 to 3 with MG80-MA with serial numbers 100001 to 299999, contact
our sales representative or service desk.