10
5. Function change when using the MG80-LM
In previous versions, measuring screen refresh was not performed while latch operation is
stopped (standby status for operation, measurement start button text color is gray). From
this version, however, normal operation is possible and the display is refreshed even while
latch operation is stopped.
When set to "Encoder"
When set to "Encoder (High-speed)"
• Measurement start/stop button
Measurement is performed using the measurement start/stop button.
The button text color changes according to the status as follows.
:
Latch operation stopped (Standby status for operation)
The button text color changes to gray after power-on, after settings are changed,
and after latch operation ends (the set latch count is reached).
Normal operation can be performed.
: Latch operation in progress, standby for encoder signal input
When the measurement start/stop button is touched while latch operation is
stopped, operation enters the standby status for encoder operation and the button
text color changes to orange.
: Latch operation in progress
When the encoder signal is detected and latch operation starts, the button text color
changes to green.
: When a latch error or encoder error occurs
When a latch timing or encoder abnormality is detected, the button text color
changes to red.
Touch the measurement start/stop button to cancel the error.
Measurement start/stop button