7. Interface Unit Settings
MODE switches
A/B signal output type
Analog output type
1, 2
POS. lamps light
MODE switch 1
MODE switch 2
MODE switch 1 ...... Always ON
OFF
OFF
MODE switch 2 ...... Always OFF
ON
OFF
OFF
ON
ON
ON
3
ON ............... B signal is leading
Always OFF
OFF .............. A signal is leading
4
Always OFF
Always OFF
5
Normally ON
Always ON
OFF only during reference point adjustment
6
Output format F: ON ..... 0.1
µ
m
Output format G: ON ...... 0.02
µ
m
Always ON
OFF .... 0.05
µ
m
OFF .... 0.01
µ
m
7
Always OFF
Always OFF
8
ON ............... Hold Mode
Always OFF
OFF .............. Automatic Reset Mode
9
Always OFF
Always OFF
10
ON ............... 1/4 Reference Point Mode
Always OFF
OFF .............. Reference Point Mode
For details about changing the settings, refer to the Instruction Manual (sold separately).
4
3
2
1
MODE switches (for various settings)
POS. lamp
4 3
2 1
1
ON
2
3
4
5
6
7
8
9
10
1
ON
Default setting
status
(In the case of
output format F, H)
Default setting
status
(In the case of
output format G)
2
3
4
5
6
7
8
9
10
(Reference point POS1 switch)
(Reference point POS2 switch)
(Direction switch)
(Always off)
(Reference point adjustment mode switch)
(Resolution setting switch)
(Always off)
(Alarm mode switch)
(Always off)
(Reference point mode switch)
MODE switch 1
MODE switch 2
MODE switch 3
MODE switch 4
MODE switch 5
MODE switch 6
MODE switch 7
MODE switch 8
MODE switch 9
MODE switch 10