20
MAGNAMED
Ventilator Analyzer -
VentMeter
P/N: 1901509
– Rev 05
Note
•
Attempting to select a seventh parameter to be shown it will not blink, because the
maximum parameters that can be shown is six.
•
If this new parameter must be shown, then first deselect one of the parameters and select
this new parameter.
•
To observe inspiratory and expiratory cycles watch the lung drawing between the F1 and
F2 key.
•
To carry out continuous pressure or flow monitoring observe that the lung between F1
and F2 key will remain static.
Configurations
(F1 key from the main screen):
By pressing F1 in the
Main Screen
it will present a screen with three flow sensors. The blinking flow sensor is
the one selected. To select another flow sensor, use the UP and DOWN keys to reach the new sensor and
press ENTER key to select it. This new sensor will start to blink indicating that it is the one selected.
Battery
(F2 key from the Configuration Screen):
By pressing F2 in the configuration screen it is possible to visualize the charging situation and the battery
temperature.
Calibrations
(F2 key from the main screen):
By pressing F2 key in the Main Screen it will present the Transducers Calibration Screen. In this screen, by
pressing ENTER key, the transducers will be calibrated. It will present the message: Calibrating… and in this
condition F1 and F2 will be inactive. After the calibration process the message will be cleared and the F1 and
F2 keys will return to the active state.