Magnacoustics Genesis ULTRA Installation Manual Download Page 5

5

 | 

Magnacoustics Inc™ © 3/2008 rev-o 

 

Technologist Control Unit Operation

 

 
 
11:  Depending on how loud the Scan Sequence is the 

AUTO GAIN

 can be turned on and off 

by depressing the Violet 

AUTO GAIN

 button (located under the Red 

MIC VOL-

 button).  

The 

AUTO GAIN

 setting can be seen on the upper left corner of the display in the 

“GAIN”

 

box (

Off, Idle, Low, Med, 

and

 High

).  When the 

AUTO GAIN

 is active you can also see an 

^ to the left of the VOL BAR indicating that it has adjusted the MUSIC LEVEL to the 
Patient.  The percentage can be adjusted to 5%, 10%, 15%, 20%, or 25% per stage in the 
PROGRAM MODE 

STEP 14. 

 
12:  Depressing the Violet 

COMP

 button (located under the Violet 

AUTO GAIN

 button) will 

provide a more even flow of Music to the Patient, by raising the lows and lowering the 
highs. 

 
13:  All the other buttons control the Stereo. 
 
14:   Depressing the Blue 

“PROGRAM”

 button (located below the red 

AVC VOL-

 button) will 

put the system in “Program Mode.  Depressing the 

MUSIC VOL+

 and 

VOL-

 button toggle 

up and down through the Menu respectively.  Depressing the Blue and Violet 

INPUT/ENTER

 button (located bellow the Blue 

PROGRAM

 button) will toggle the options 

in the menu that is highlighted.  Depressing the 

PROGRAM 

button again will memorize any 

changes you have made and exit the Program Mode. (use Page 6 for visual reference of 
button locations)  

 

BOLD

 text indicates the Default Settings. 

 

Tree of programming options: 

 

 

1.Equalizer: 

YOKE 1

 

YOKE 2 

MAGNAMUFF 

MAGNAPLUG 

FLAT 

 

2. Gain Increment: 

5% 

10% 

15% 

20%

25%  

3. Gain Dwell Time: 

2Sec

 

4Sec 6Sec 

 

 

 

4. Initialization: 

TUNER

 

VIDEO FMRI 

XM 

iPod 

CD 

5. Volume Norm Set: 

20

 

30 

40 50  

 

6. Volume Gain Set: 

LOW 

NORM

HIGH  

   

7. Bass: 

LOWEST 

LOW 

NORM

HIGH HIGHEST 

 

8. Treble: 

LOWEST 

LOW 

NORM

HIGH HIGHEST 

 

9. Contrast: 

MIN 

2, 3,4,5,6 

7,

  NORM

 

9,10,11, 12,13,14, 

MAX 

 

10. Keypad Click: 

NORM

 

OFF 

   

 

11. Ambient MIC Level  Set 

 

 

 

 

 

12. Scan MIC Level 

Set 

 

 

 

 

 

13. Boot Mode 

Norm

 

Diagnostic 

   

 

14. FMRI 

Pulse High

  Pulse Low 

High Held 

Low Held 

 

 

15. IR Remote 

JVC

 

 

 

 

 

 

16. IR Remote Dwell 

1Sec

 

2Sec 3Sec 

4Sec 

   

 

 
 

   

 

 

   

 

 

 

 

   

 

 

 

 

   

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Summary of Contents for Genesis ULTRA

Page 1: ...llation Manual Index System Notes 2 Headset Operation 3 Technologist Control Unit Operation 4 5 and 6 Important Safety Instructions 7 Pre Installation 8 System Parts List 9 10 Genesis ULTRA Installati...

Page 2: ...hanges volume automatically to mask MR gradient sounds Patient volume and music selection controls with voice feedback for maximum comfort and versatility Backlit Technologist Control Unit allows oper...

Page 3: ...to provide the highest level of attenuation it is necessary to first rotate tops of Headset toward the rear of the patients head then while applying pressure rotate fully forward to align with the ear...

Page 4: ...roviding the intercom integration kit was installed the Display will show AVC in the lower left MODE box and the music to the patient will be interrupted The AVC level and the console intercom level w...

Page 5: ...C VOL button will put the system in Program Mode Depressing the MUSIC VOL and VOL button toggle up and down through the Menu respectively Depressing the Blue and Violet INPUT ENTER button located bell...

Page 6: ...6 Magnacoustics Inc 3 2008 rev o...

Page 7: ...g periods of time 14 Refer all servicing to qualified service personnel Servicing is required when the apparatus has been damaged in any way such as power supply cord or plug is damaged liquid has bee...

Page 8: ...ck for the following items 1 AM FM Multi Disc CD Changer 2 One AM antenna 3 One FM antenna 4 Two Speakers 5 One Transducer Module houses connections for Headsets Patient Volume control Patient selecti...

Page 9: ...s over IR receiver on the Stereo 18 CB8 6ft AC Adapter Cable connects the AC Outlet to the Switched Voltage Selectable and Fused AC Receptacle on the rear of Genesis module 19 AutoVoice Breakout Box i...

Page 10: ...minals on the back of the Stereo Red and Black to Left Red and Black Green and White to Right Black and Red respectively The other side of cable CB6 to the upper terminals of SPK1 on the rear of Genes...

Page 11: ...Cable install a 75 to 300 ohm adapter to eliminate any loops and only connect 1 wire 16 For GE MRI s remove block off plate from J53 or J52 at penetration panel all others use supplied kit or fabricat...

Page 12: ...rcuit B Power If not pull AC Power and contact Magnacoustics Tech Support at 1 800 637 2282 or E mail at support magnacoustics com for further details 36 Press the green Power Button on the TCU a self...

Page 13: ...8 Tech Unit Connector J2 fMRI Input J9 RS232 Spare J3 Video Input J10 RS 232 Spare J4 MIC Input J11 To Transducer CB1 J5 AVC Input J12 AC Receptacle 5 amp SB J6 fMRI trigger output SPK1 Speaker Juncti...

Page 14: ...eo Multi CD Player SCAN ROOM MRI Pen Panel Pneumatic Transducer Patient Headset Pneumatic Tubing IR Cable CB 5 CB 2 50 Technologist Control Unit Patient Mic Patient Controls DB9 Filter Gradient Mic Sp...

Reviews: