Magnacoustics Genesis ULTRA Installation Manual Download Page 11

11

 | 

Magnacoustics Inc™ © 3/2008 rev-o 

 

Genesis ULTRA Installation and Connections continued

 

 
13: 

A.  For Horizon LX remove run 788 from J7 on the WIM Module and connect to the open 

connector on the AutoVoice Breakout Box, then connect the other end of “CB4” to J7 on the 
WIM.  For 5X Console remove the cable going to J9 on the Auto Voice Board and connect to 
the open connector on the AutoVoice Breakout Box, then connect the other end of “CB4” to J9 
on the Auto Voice Board. 

B. For 

Excite systems 

remove Run 

1085

 labeled 

OW1-A21-J7

 from 

J7

 on the Wim Module and 

connect to the open connector on the AutoVoice Breakout Box, then connect the other end of 
“CB4” to 

J7

 on the Wim.  For 5X console remove the cable going to 

J9

 on the Auto Voice 

Board and connect to the open connector on the AutoVoice Breakout Box, then connect the 
other end of “CB4” to 

J9

 on the Auto Voice Board. 

 
14:  Connect the receptacle side of cable “CB8” 6ft AC Adapter Cable to 

J12 

on the rear of Genesis 

module and 

DO NOT PLUG INTO AC OUTLET UNTILL ALL CONECTIONS ARE MADE. Also 

ensure that the power entry receptacle is set up for your proper voltage 110 / 220.   

15:  Connect FM and AM Antenna to the Stereo 

Reference Antenna Connections in the Stereo 

Instructions for proper installation of FM / AM Antenna Do Not Connect any external 
Antenna Grounds to the Stereo (A ground loop will be created and you will hear a 60 Hz 
Hum) If using building Antenna or Cable install a 75 to 300 ohm adapter to eliminate any 
loops and only connect 1 wire 

16:  For GE MRI’s remove block off plate from J53 or J52 at penetration panel, all others use 

supplied kit or fabricate a DB9 slot for filter as required.  

17:   Install RF filter with 

female toward computer room. 

18:  Locate existing electrical conduit between control console and computer room. 

 

CAUTION

 

Assistance may be necessary to ensure Trauma free installation of cable CB1 

 
19:  Route and install Cable “CB1” (90 ft. DB9 Male to Female) with Female connector toward 

Genesis Module and Male toward penetration panel. 

20:   Connect “CB1” Male to the RF filter  
 
21:  Zip tie cable “CB1” to existing supports, or cables from penetration panel to provide strain relief. 
22:  Connect Female end of cable “CB1” to 

J11

 on the rear of Genesis module.

 

 

23:   From scan room side of penetration panel connect either end of cable “CB2” (50ft. DB9 Female 

to Female) to the RF filter you installed on penetration panel. 

 
24:  Zip tie cable “CB2” to existing cables from penetration panel to provide strain relief. 
 
25.  Route cable “CB2” on opposite side of the cold head (if used) to the front of the MRI utilizing 

existing conduits. 

 
26:  Leave approximately 18 in. of cable “CB2” protruding from the front center of the MRI directly 

above the table docking mechanism, or alongside the base of the gantry.  It is important to 
locate the end of “CB2” as close as possible to the front center of the bore.  

 
27:  Connect cable “CB2” to the Transducer and place it on the pedestal above the height control 

peddles, or locate the Transducer as close as possible to the front center of the bore. 

 

Summary of Contents for Genesis ULTRA

Page 1: ...llation Manual Index System Notes 2 Headset Operation 3 Technologist Control Unit Operation 4 5 and 6 Important Safety Instructions 7 Pre Installation 8 System Parts List 9 10 Genesis ULTRA Installati...

Page 2: ...hanges volume automatically to mask MR gradient sounds Patient volume and music selection controls with voice feedback for maximum comfort and versatility Backlit Technologist Control Unit allows oper...

Page 3: ...to provide the highest level of attenuation it is necessary to first rotate tops of Headset toward the rear of the patients head then while applying pressure rotate fully forward to align with the ear...

Page 4: ...roviding the intercom integration kit was installed the Display will show AVC in the lower left MODE box and the music to the patient will be interrupted The AVC level and the console intercom level w...

Page 5: ...C VOL button will put the system in Program Mode Depressing the MUSIC VOL and VOL button toggle up and down through the Menu respectively Depressing the Blue and Violet INPUT ENTER button located bell...

Page 6: ...6 Magnacoustics Inc 3 2008 rev o...

Page 7: ...g periods of time 14 Refer all servicing to qualified service personnel Servicing is required when the apparatus has been damaged in any way such as power supply cord or plug is damaged liquid has bee...

Page 8: ...ck for the following items 1 AM FM Multi Disc CD Changer 2 One AM antenna 3 One FM antenna 4 Two Speakers 5 One Transducer Module houses connections for Headsets Patient Volume control Patient selecti...

Page 9: ...s over IR receiver on the Stereo 18 CB8 6ft AC Adapter Cable connects the AC Outlet to the Switched Voltage Selectable and Fused AC Receptacle on the rear of Genesis module 19 AutoVoice Breakout Box i...

Page 10: ...minals on the back of the Stereo Red and Black to Left Red and Black Green and White to Right Black and Red respectively The other side of cable CB6 to the upper terminals of SPK1 on the rear of Genes...

Page 11: ...Cable install a 75 to 300 ohm adapter to eliminate any loops and only connect 1 wire 16 For GE MRI s remove block off plate from J53 or J52 at penetration panel all others use supplied kit or fabricat...

Page 12: ...rcuit B Power If not pull AC Power and contact Magnacoustics Tech Support at 1 800 637 2282 or E mail at support magnacoustics com for further details 36 Press the green Power Button on the TCU a self...

Page 13: ...8 Tech Unit Connector J2 fMRI Input J9 RS232 Spare J3 Video Input J10 RS 232 Spare J4 MIC Input J11 To Transducer CB1 J5 AVC Input J12 AC Receptacle 5 amp SB J6 fMRI trigger output SPK1 Speaker Juncti...

Page 14: ...eo Multi CD Player SCAN ROOM MRI Pen Panel Pneumatic Transducer Patient Headset Pneumatic Tubing IR Cable CB 5 CB 2 50 Technologist Control Unit Patient Mic Patient Controls DB9 Filter Gradient Mic Sp...

Reviews: