Performance adjustments 

 
Performance  adjustments  to  your  power  chair  should  only  be  made  by  professionals  of  the 
healthcare  field,  or  by  persons  fully  conversant  with  both  this  process  and  the  driver’s 
capabilities.  
 

 

WARNING!  Changing  the  performance  settings  could  adversely  affect  your 
power chair. You may cause injury to yourself and to others. 

 

STOP!  Incorrect  settings  could  cause  damage  to  the  chair  and  to  surrounding 
property. 

 

5.2 

Positioning 

 
If  your  power  chair  was  configured  at  your  Magic  Mobility  Dealer,  please  consult  your  health 
care  professional  before  changing  the  seat  position  or  making  any  other  adjustment.  Some 
adjustments may degrade your power chair’s performance and safety by changing its centre of 
gravity.  
 

5.3 

Freewheel Mode 

– pushing the power chair 

 

The  Extreme  X8  is  equipped  with  a  manual  freewheel  system  enabling  the  electromagnetic 
brakes to be disengaged and the power wheelchair to be manoeuvred manually. To disengage 
the built-in brakes simply push the two levers downwards as shown in 

Figure 4

 
The wheelchair controls will not function and the “status” light on the joystick will flash when the 
chair is in free-wheel mode. This feature has been intentionally incorporated to protect the user 
from unsafe situations. 
 
Don’t forget to push the levers back in again firmly after manually positioning the chair. 

 

 

WARNING!  Do  not  use  your  chair  in  freewheel  mode  or  attempt  to  place  your 
chair into freewheel mode without an attendant present. You may cause injury to 
yourself and to  others. Do  not  place  your  power  chair  in freewheel  mode  while 
on an incline. The chair could roll uncontrollably on its own, leading to injury to 
yourself and others.  
When  the  power  wheelchair  is  in  freewheel  mode,  the  braking  system  is 
disengaged. 

 

5.4 

Electrical Safety Protection 

 

Your  wheelchair  has  a  fuse  fitted  into  the  battery  circuit  to  protect  the  motors  and  electronics 
from an accidental current overload. In the case of a blown fuse, the chair will not drive and you 
will need to contact your Magic Mobility dealer for repair / replacement 

 

5.5 

Rearward Stability Wheels 

 

Summary of Contents for Extreme X8 with G90

Page 1: ...Page 1 Owner s Manual X8 with G90...

Page 2: ...Page 2 www magicmobility com au YouTube MAGIC MOBILITY OZ...

Page 3: ...testing 12 5 Operating Instructions 12 5 1 Performance adjustments 13 5 2 Positioning 13 5 3 Freewheel Mode pushing the power chair 13 5 4 Electrical Safety Protection 13 5 5 Rearward Stability Wheel...

Page 4: ...ge 25 8 8 Transportation 26 8 9 Daily Checks 26 8 10 Weekly Checks 26 8 11 Annual Checks 26 9 Servicing 26 10 Fault finding 27 10 1 Limp Mode 27 10 2 Stuck Power Button 27 10 3 Joystick out of neutral...

Page 5: ...t the time of printing However due to our policy of continual product improvements we reserve the right to make changes at any time without notice This may lead to slight variations between the illust...

Page 6: ...f this manual are based on the expectation that a qualified healthcare professional has properly fitted the power chair to the user and the prescribing healthcare professional has trained the user in...

Page 7: ...chair SEATING ARRANGEMENT MAXIMUM USER WEIGHT No seat elevator fitted 182 kg Seat elevator or tilt fitted 155 kg STOP Exceeding the weight capacity voids your warranty and may result in damage to your...

Page 8: ...pable of climbing an obstacle of up to 100mm in height however this can be much lower depending on the setup of the wheelchair Never try to travel backwards down any step kerb or other obstacle This m...

Page 9: ...bility recommends that the maximum slope of an incline you attempt to safely ascend or descend on your power chair does not exceed a 1 in 14 gradient Given the off road capabilities of the power chair...

Page 10: ...electromagnetic fields This will eliminate the possibility of unintended movement caused by electromagnetic sources Failure to take this precaution may result in personal injury 3 12 Positioning Belt...

Page 11: ...not operate your power chair while you are under the influence of alcohol as this may impair your ability to operate your power chair in a safe manner resulting in personal injury to yourself and oth...

Page 12: ...tained from the power chair manufacturer If you would like access to this information please contact your Magic Mobility Dealer 5 Operating Instructions The speed and direction of the power chair is c...

Page 13: ...r to be manoeuvred manually To disengage the built in brakes simply push the two levers downwards as shown in Figure 4 The wheelchair controls will not function and the status light on the joystick wi...

Page 14: ...s upright position When returning to the upright position always be sure that the mechanism has reached its lowest limit or normal seated position WARNING Never tilt the seat from its upright position...

Page 15: ...r operator via the joystick Bring your power chair to a complete stop on a flat level surface Push the Accessory Mode Selection button on the DX2 joystick Select the Steering lock option by moving the...

Page 16: ...joystick module is splash proof but may be permanently damaged if water transgresses the rubber seals this damage is not covered under warranty We recommend carrying a plastic bag large enough to cov...

Page 17: ...Status LED will illuminate and not flash Press the On Off button again to turn the power off 6 5 2 Right Indicator If Lights Are Fitted To turn the Right Indicator on off press and release short press...

Page 18: ...file will increase by a single increment with each press until the maximum configured value is reached The Mode button does not enter the accessory modes 6 5 6 ASK Accessory Shortcut Key If you are in...

Page 19: ...ill beep and automatically power itself down To Unlock the system Press the On Off button to power up the system The flashing red key symbol shows the system is locked Swipe the magnetic key across th...

Page 20: ...WARNING When fitting alternate batteries ensure their terminal posts cannot touch any part of the wheelchair frame STOP Automotive batteries and chargers are unsafe for use in power wheelchairs 7 2 B...

Page 21: ...ip Avoid stop go driving try to maintain a constant speed Try to avoid inclines Limit baggage weight carried 7 4 Public Transportation The Gel Cell batteries are Federal Aviation Administration FAA ap...

Page 22: ...so it may be pinched in any of the seat or power base frames Some parts of the power wheelchair are susceptible to extremes of temperature It is possible for the batteries to freeze The temperature at...

Page 23: ...ly All pneumatic tyres are fitted with automotive type valves and can be inflated using most typical automotive hand and foot type pumps as well as service station air outlets STOP Under inflated tyre...

Page 24: ...o resist corrosion Although all effort has been made to ensure the long term durability of the product we cannot guarantee that the wheelchair will remain corrosion free for the duration of its usable...

Page 25: ...hosed down The wheelchair should then be left in a warm and dry environment It may also help to blow as much of the sand and or salt off as possible 8 6 5 Preventative Maintenance Prior to using your...

Page 26: ...tick slowly forward until you hear the electric brakes click Immediately release the joystick The brakes should operate within a few seconds of moving the joystick Repeat this test by pushing the joys...

Page 27: ...remanufacturing and may render your chair unsafe 10 Fault finding During fault conditions the chair either will not drive or drives very slowly 10 1 Limp Mode If the DX system detects a fault that doe...

Page 28: ...g Seat elevator or tilt fitted 155 kg Maximum Speed 10km h 6 2mph Estimated Range 25km 15 5 miles 73Ah gel cell batteries dependent on wheelchair setup and local conditions Turning Radius 2000mm 79 Gr...

Page 29: ...s is subject to the discretion of the provider not the manufacturer For more information regarding replacement units contact your Magic Mobility Dealer Upon acceptance of goods at delivery the purchas...

Page 30: ...ffice and Operations 2 112 Browns Road Noble Park Vic 3174 Australia Phone 61 3 8791 5600 Fax 61 3 9701 0144 Email Address admin magicmobility com au Website http www magicmobility com au Copyright Ma...

Page 31: ...eat and use the vehicle installed occupant restraint system when this is feasible When seat positioning options are fitted the following should be followed unless necessary for postural or medical req...

Page 32: ...ble crashworthy wheelchair anchored pelvic belt restraint when WC4 19 ISO7176 19 tie downs are requested This belt is designed to accommodate use on either side of the vehicle The belt is equipped wit...

Page 33: ...he junction of the shoulder belt and pelvic belt of three point belts should be located near the hip opposite to the shoulder over which the diagonal belt crosses and not near the midline of the occup...

Page 34: ...mid size adult male weighs between 75 and 136Kgs and the appropriate ATD crash test dummy would weigh approximately 76Kg This weelchair was tested with a 76Kg ATD Occupants with a higher weight are at...

Page 35: ...own and occupant restraint system and the manufacturer s instructions Whenever possible auxiliary wheelchair equipment should be removed from the wheelchair and stored in a cargo area or secured in th...

Page 36: ...tacted by a wheelchair occupant s head during a side impact collision or a vehicle rollover should be padded with material that complies to FMVSS 201 Only spill proof batteries e g gel cell should be...

Page 37: ...sts The wheelchair should be inspected by a manufacturer s representative before reuse following involvement in any type of collision Care should be taken when applying the occupant restraint to posit...

Page 38: ...Page 38...

Page 39: ...unit Examples include citizens band CB radios walkie talkie security fire and police transceivers cellular telephones and other personal communication devices NOTE Some cellular telephones and simila...

Page 40: ...d wheelchair may make it more susceptible to EMI Note There is no easy way to evaluate their effect on the overall immunity of the powered wheelchair Report all incidents of unintended movement or bra...

Reviews: