Troubleshoots
Trouble
Possible Causes
Recommendations
Liquid dropped on Cartridge
carrier or Tip carrier
Suction leakage, due to damaged or worn out
O-rings
Please refer to system test and check if the liquid level is the
same throughout tips.
Wrong liquid quantity moved
1. Suction head might be bent.
2. Worn out O-ring
3. Reagent is stored for too long.
1. Check suction head.
2. Please refer to system test and check if the liquid level is
the same throughout tips.
3. Renew your reagents.
Liquid level floods over tip filter
1. Wrong sample materials.
2. Wrong sample tubes.
1. Please refer to MagCore® kit manual.
2. Please use the provided sample tubes.
No piercing
V motor or connector is not working normally. Restart the system. Please contact technical support if the
problem persists.
No axis movement
Motor or connector is not working normally.
Restart the system. Please contact technical support if the
problem persists.