18

MagCore® HF48

Select Moving Test in Maintenance menu.

Select the axis to test and press Start. The motor will run 
continuously until you press Stop.

Select DC Motor Operate in Maintenance menu.

 All consumables are fixed in place when test is stopped. You can 
operate the DC motor through this function if you need to raise or 
lower the drip pan.
Refer to the page 7 for corresponding locations of operation areas.

Select the area you want to move up or down and press Start. It will 
stop automatically when the motor moves to end.

━ Moving Test

━ DC Motor Operate

Next Step

Next Step

X axis motor running test

Operation Area A UP

Shutter Open

Operation Area B UP

Y axis motor running test

 Operation Area A DOWN

Shutter Close

Operation Area B DOWN

M axis motor running test

V axis motor running test

Moving tests for all axes 

Summary of Contents for HF48

Page 1: ...tallCartridge 7 TouchPanelandUserInterface 8 Main Menu 8 PanelFunctionalArea 8 EmergencyStopandWarning 9 Start 10 LoadProgram 11 SavePrograms 12 DeletePrograms 12 Update Program Files 13 EasyMaintenan...

Page 2: ...erSources Thisproductshouldbeoperatedonlyfromthetypeofpower sourceindicatedonthelabel Ifyouarenotsureofthetypeofpowersupplyat your lab consulttheseller or local power company For products intended to...

Page 3: ...ationParameters OperationEnvironment Applications Whole Blood DNA Viral Nucleic Acid DNA RNA 200 l 400 l 1200 l Plasma and Circulating Cell Free DNA 1200 l Tissue Genomic DNA Plant Genomic DNA Bacteri...

Page 4: ...Tip TipHolderSet X50 ElutionCarrier X1 CartridgeCarrierB X1 CartridgeCarrier A X1 Tip Carrier B X 1 TipCarrierA X 1 PowerCord X 1 Lubricant X1 O Ring X24 Systemtesttoolsshouldbekeptincleanplacesandth...

Page 5: ...Shutter Leveling Glides four legs Fixed Knob on the Top Cover of Part Chamber SystemOverview Cooling Fans Power Switch and Plug Power Switch red Light Switch green Part Chamber Maintenance Cover Stat...

Page 6: ...on close to outlets and leave sufficient space for system installation and heat dissipation Pleaesuse theleveling glidestofine tune theelevationlevel System Height 160 cm System Depth 80 cm System Wid...

Page 7: ...etheelevation Remembertotightenthescrews Cutthefixationbeltandclosethetopcoverofpartchamber Cutthefixationbeltsontheracks Pluginthepowercordandturnonthemainpower Turnonthepowerandthesystemwillstartini...

Page 8: ...areaA Carrierofarea B canonly be used in area B Incorrect usemay causesystem error Install Cartridge Install Tipand SampleTube InsertTipsandSampleTubesintotheslotsintheTipCarrier Then placetheTipCarr...

Page 9: ...m Load frequently used programs stored by user Maintenanceandothertestfunctions Basicmaintenance UVsterilization and systemtests Systemtime Systemstatus Basicsysteminformation serialnumber errormessag...

Page 10: ...2 Resume Thesystemwillstopalloperations immediatelyifthefrontdoorisopened SelectContinuetoresumetoprevious operationsorAborttoendalloperations PressStopkeytostopalloperationsimmediately incaseofemerge...

Page 11: ...ratethesystemifthe parametersarecorrect PleasepressNotomodifyincorrect parameters Youmaymonitortheextractionprogressandchecktheremaining timeonthescreen ThesystemwilldisplayCompletepageafter completio...

Page 12: ...arecorrectandpressYes Please select Cartridge Code Please select Channel NextStep NextStep Load Program Youcansavefrequentlyusedprogramswithnamesyoupreferforyour convenience PressLoadProgramtoenterthe...

Page 13: ...ktomodify Confirmiftheparametersarecorrect EnterthenameyouwanttosaveandpressEnter SavePrograms Please select Channel Youmaydeleteyoursavedprogramaswell PleasepressLoadProgramstoenterStoredProgramspage...

Page 14: ...neby users Theseitemsincludesterilization cleaning systemtests errormessage report etc Usersmustperformtheseitemsregularlytoensurethatthesystem canfunctionnormally UpdateProgram Files UV Sterilization...

Page 15: ...rings PressFinishafterapplyinglubricant ReplaceO Rings SelectCleanPiercingNeedlesandpressSTART Pleaseprepareacottonpaperwith70 alcoholtoclean Clean PiercingNeedles NextStep NextStep NextStep NextStep...

Page 16: ...ipssmoothly Ifyes pressNext Ifyou findanytipstuckinitstipholder pressAborttostopthetestand replacethetip tipholder Runthetestagainandiftheproblem persists contactyourlocaldistributorfortechnicalsuppor...

Page 17: ...mparewiththevalueonyourthermometer toseeiftheyarethesame PleasepressStopifyouwanttostopthetest HeaterTest Thisfunctionisforthesystemtoreturntipstotheiroriginal locationsafteremergencystoporoperationin...

Page 18: ...ostopifyouwanttostopsterilization UV Sterilization OpenthetopcoverofthePartChamber ReplacetheoldUVlampwithanewone SelectStatusinthemainmenuandpressUVTimeResetkeyto resetthehoursofUVlamp Pleasetesttose...

Page 19: ...sfunctionifyouneedtoraiseor lowerthedrippan Refertothepage7forcorrespondinglocationsofoperationareas SelecttheareayouwanttomoveupordownandpressStart Itwill stopautomaticallywhenthemotormovestoend Movi...

Page 20: ...ce com Selecttheitemsyouwishtochange EntersettingvalueandpressEnter SystemTiming Setting YearSetting MonthSetting DaySetting 12or24hourdisplayformat MinuteSetting HourSetting NextStep NextStep Setup Y...

Page 21: ...videthefilefor engineerstodetermineandsolveerrorsinthesystem Systemdate Systemtime Machineserialnumber PLCversion UIversion UVlampoperatingtime UVlampoperatingtime Errormessagelistoutput NextStep Next...

Page 22: ...ied for replacement Using fuses with higher amps maycausedangerofburning ReplaceFuses Others Youcanusethebarcodescannertoinputyoursample scodeinto thetestreport Select BarcodeScan functionandenter Bar...

Page 23: ...rtoolong 1 Checksuctionhead 2 Please refer to system test and check if the liquid level is thesamethroughouttips 3 Renewyourreagents Liquidlevelfloodsovertipfilter 1 Wrongsamplematerials 2 Wrongsample...