www.madur.com
Manual
GA-21Plus
-
in the
GAS
line. Confirm standard gas editing. Sup-
ply the instrument with standard gas for about 4 min.
-
in the
STORE
line, as soon as the value has stabil-
ised, stores the measured value as calibration value.
Disconnect the standard gas and flood the instru-
ment for at least 15 min with ambient air.
-
return to select the next gas, or end the calibration
5.6.2.6. Calibration adjustment
Calibration adjustment is a new feature (introduced in firmware version 2.07). Al-
lows to perform a quick correction to a factory calibration using one calibration mixture
(mixture of gases in N2 balance).
Calibration adjustment does not substitute factory calibration with a single gas in
N2 balance. I
.e.
calibration adjustment cannot create matrix of cross-sensitity signals –
this is created only during standard calibration.
- means that correction for a selected gas is applied.
If all sensor are unmarked: , it means no adjustment to factory calibration is ap-
plied.
To perform calibration adjustment:
Enter the concentration for each gas present in mixture. If any of listed gas is not
present in mixture set its concentration to 0 → analyser will not include this sensor in ad-
justment.
42