![Madur GA-21 Plus Operating Manual Download Page 22](http://html.mh-extra.com/html/madur/ga-21-plus/ga-21-plus_operating-manual_688625022.webp)
www.madur.com
Manual
GA-21Plus
The analyser does not switch to
STANDBY
mode if [
OFF
] is pressed in case:
-the key is being pressed during Initial Calibration. In that case the instru
ment realises the action, but will switch to
STANDBY
only after the cal
ibration is finished.
-the key is being pressed while the is still a high gas concentration detec
ted in the gas chamber. In this case the instrument will automatically
switch to
FLOODING WITH AMBIENT AIR
mode and switch to
STANDBY
mode later on.
The following screen is shown in the
STANDBY
mode.
By pressing [
OFF
] the instrument can be switched off completely. If [
C
] is
pressed the instrument switches ON again. If the instrument is in
STANDBY
mode for
more than 8.5 hours it switches off automatically. It is highly recommended to carry out a
calibration of the oxygen
sensor when re-activating the instrument.
22
Summary of Contents for GA-21 Plus
Page 1: ...FLUE GAS ANALYSER GA 21 plus Operating manual Version 2 1 06 2017 ...
Page 15: ...www madur com Manual GA 21Plus Picture 2 Front elevation picture 15 ...
Page 44: ...www madur com Manual GA 21Plus 44 ...
Page 52: ...www madur com Manual GA 21Plus 52 ...