Specifications

Specifications are subject to change without notice. Specific warranty remedy limitations apply. Call (603) 456-2011 or go t

madgetech.com

 for details.

TEMPERATURE

Temperature Range

-20 °C to +60 °C (-4 °F to +140 °F)

Resolution

0.01 °C (0.018 °F)

Calibrated Accuracy

±0.50 °C/± 0.18 °F

(0 °C to +55 °C/32 °F to 131 °F)

Response Time

10 minutes free air

REMOTE CHANNEL

Thermocouple 

Connection

Female subminiature (SMP) (MP model)

Pluggable screw terminal (TB model)

Cold Junction

Compensation

Automatic, based on internal channel

Max. Thermocouple

Resistance

100 Ω

Thermocouple K

Range (°C): 

-65 °C to +40 °C

Resolution: 

0.1 °C

Accuracy: 

±0.5 °C

Response Time

τ

 = 2 minutes to 63% of change

GENERAL

Reading Rate

1 reading every second up to 1 reading every 24 

hours

Memory

16,128 readings

LED 

Functionality

3 status LEDs

Wrap Around

Yes

Start Modes

Immediate and delay start

Calibration

Digital calibration through software

Calibration Date

Automatically recorded within device

Battery Type

9 V lithium battery included;

user replaceable with any 9 V battery

Battery Life

3 years typical at 1 minute reading rate

Data Format

For Display:

 °C or °F

For Software:

 Date and time stamped °C, K, °F or °R

Time Accuracy

± 1 minute/month

Computer 

Interface

USB to mini USB, 250,000 baud for standalone 

operation

Operating 

System 

Compatibility

Windows XP SP3 or later

Software 

Compatibility

Standard Software version 4.2.21.0 or later

Operating 

Environment

-20 °C to +60 °C (-4 °F to +140 °F),

0 %RH to 95 %RH non-condensing

Dimensions

3.0 in x 3.5 in x 0.95 in 

(76.2 mm x 88.9 mm x 24.1 mm) Data logger only 

Glycol Bottle

30 mL

Probe Length

72 in

Material

ABS Plastic

Weight

4.5 oz (129 g)

Approvals

CE

Alarm

User configurable high and low audible and on-

screen alarms. 
Alarm Delay: A cumulative alarm delay may be 

set in which the device will activate the alarm (via 

LED) only when the device has recorded a user 

specified time duration of data.

Audible Alarm 

Functionality

1 Beep per second for reading alarm above/

below threshold

BATTERY WARNING: 

BATTERY MAY LEAK, FLAME OR EXPLODE IF DISASSEMBLED, SHORTED, CHARGED, CONNECTED TOGETHER, MIXED WITH USED OR OTHER BATTERIES, EXPOSED 

TO FIRE OR HIGH TEMPERATURE. DISCARD USED BATTERY PROMPTLY. KEEP OUT OF REACH OF CHILDREN.

Ordering Information

VFC2000

PN 902238-00

VFC Temperature data logger with USB to mini USB cable

Power Adapter

PN 901839-00

Replacement USB universal power adapter

U9VL-J

PN 901804-00

Replacement battery for VFC2000