MadgeTech VFC2000 Product User Manual Download Page 2

 Product User Guide  |  2

VFC2000

PRODUCT USER GUIDE

Product Overview

The VFC2000 is a simple 
solution for vaccine 
temperature monitoring 
compliance. Designed to 
satisfy all CDC and VFC 
requirements, the VFC2000 
provides highly precise, 
continuous temperature 
validation. Featuring a 
convenient LCD screen, the 
VFC2000 displays current 
readings, minimum and 
maximum statistics as well 
as a battery level indicator. 
User-programmable 
alarms trigger an audible 
and visual alert. Optional 
AC power source allows 
battery to be a back up if 
there is a power loss. 

VFC Requirements

•  Detachable, buffered temperature probe
•  Out-of-range audible and visual alarms
•  Low battery indicator with external power and battery 

backup

•  Current, minimum, and maximum temperature display
•  Accuracy of ±0.5°C (±1.0°F)
•  Programmable logging interval (1 reading per second 

to 1 reading per day)

•  Daily check reminder warning
•  Suitable for vaccine transportation
•  Also monitors ambient room temperature

Device Operation

1.  Install the MadgeTech 4 Software onto a Windows PC.
2.  Connect the data logger to the Windows PC with the 

provided USB cable.

3.  Launch the MadgeTech 4 Software. The VFC2000 will 

appear in the Connected Devices window indicating 
the device has been recognized.

4.  Select the start method, reading rate and any other 

parameters appropriate for the desired data logging 
application. Once configured, deploy the data logger 
by clicking the Start icon.

5.  To download data, select the device in the list, click the 

Stop icon, and then click the Download icon. A graph 
will automatically display the data.

Selection Buttons

The VFC2000 is designed with three selection buttons:

Scroll:

 Allows user to scroll through 

current readings, average stats, daily 
minimum and maximum temperatures 
and device status information displayed 
on the  LCD screen.

°C

°F

Units: 

Allows users to change displayed 

units of measurement to either Celsius or 
Fahrenheit.

Start/Stop: 

To activate Manual Start, 

arm the device through the MadgeTech 4 
Software. Hold the button for 3 seconds. 
There will be two beeps confirming the 
device is started.  Reading will be displayed 
on the screen and status in software will 
change from 

Waiting to Start

 to 

Running

To pause logging while running, hold 
button for 3 seconds. 